Check the autoextraction status and manually re-extract
After a run is complete, the data from each capillary is autoextracted from the
database and a data file is created (AB1 files for sequence analysis, FSA files for
fragment analysis).
Samples can be re-extracted with the same settings, or with different analysis
protocols or results group.
You can re-extract data for several reasons.
• To extract data and create data files if autoextraction did not occur.
Example: If a network destination location that is specified in the results group is
not available, data is not autoextracted and data files are not created.
• To save data files with different names or in different locations than autoextracted
files (extract with a different results group).
• To save data files with different analysis settings than autoextracted files.
1.
In the navigation pane of the Data Collection Software, select GA
Instruments
4
ga3730
4
Run History.
2.
Search for the plate record of interest. See “Search for a plate record“ on
page 77.
All completed runs from the plate are listed and can be reextracted. Pending runs
from the plate are not listed.
3.
Select a run.
4.
Click
(Reextraction) in the navigation pane.
The Reextraction window opens. Click-drag the Result column to display
messages.
Autoextraction
and re-extraction
Re-extract runs
Chapter 5
Run the instrument
Check the autoextraction status and manually re-extract
5
72
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide