![Applent Instruments AT526 Скачать руководство пользователя страница 35](http://html.mh-extra.com/html/applent-instruments/at526/at526_user-manual_2954887035.webp)
35
Command Reference
SEND> COMP:TOL:RNOM 1E-6
<NL>
Query Syntax
COMParator:TOLerence:RNOMinal?
Query Response
<scifloat>
Example
SEND> COMP:TOL:RNOM?
<NL>
RET> +1.00000e-03
<NL>
9.8.5
COMParator:TOLerence:RLiMiT
The COMParator:TOLerence:RLiMiT command sets resistance lower/upper limit
values.
Command Syntax
COMParator:TOLerence:RLiMiT
<lower>,<upper limit
>
Parameter
Where
,<lower>,<upper
> is:
lower <float> lower limit value
upper
<float> upper limit value
Example
SEND> COMP:TOL:RLMT 100m,330m
<NL>
Query Syntax
COMParator:TOL:RLMT? <n>
Query Response
<float: lower limit>,<float: upper limit>
Example
SEND> COMP:TOL:RLMT?
<NL>
RET> 1.000000e-01,3.300000e-01
<NL>
9.8.6
COMParator:TOLerence:VNOMinal
The COMParator:TOLerence:VNOMinal command sets resistance nominal value for
the tolerance mode of the comparator function.
Command Syntax
COMParator:TOLerence:RNOMinal <
float
>
Example
SEND> COMP:TOL:RNOM 100m
<NL>
SEND> COMP:TOL:RNOM 1E-6
<NL>
Query Syntax
COMParator:TOLerence:RNOMinal?
Query Response
<scifloat>
Example
SEND> COMP:TOL:RNOM?
<NL>
RET> +1.00000e-03
<NL>
9.8.7
COMParator:TOLerence:VLiMiT
The COMParator:TOLerence:VLiMiT command sets voltage lower/upper limit values.
Command Syntax
COMParator:TOLerence:VLiMiT
<lower>,<upper limit
>
Parameter
Where
,<lower>,<upper
> is:
lower <float> lower limit value
upper
<float> upper limit value
Example
SEND> COMP:TOL:VLMT 1.1,3.9
<NL>
Query Syntax
COMParator:TOL:VLMT? <n>
Query Response
<float: lower limit>,<float: upper limit>
Example
SEND> COMP:TOL:VLMT?
<NL>
RET> 1.00,3.00
<NL>
9.9
TRIGger
Subsystem
The TRIGger subsystem command group is used to enable a measurement and to set the
trigger mode.
Figure 9‐5
TRIGger Subsystem Command Tree
TRIGger
[:IMMediate]
:SOURce
{INT,MAN,EXT,BUS}
TRG
9.9.1
TRIGger[:IMMediate]
The TRIGger[:IMMediate] command causes the trigger to execute one measurement
under BUS trigger mode.
Command Syntax
TRIGger[:IMMediate]
Example
SEND> TRIG
<NL>
Note