[SETUP] Key
5.
[SETUP]
Key
This chapter describes:
<STEUP> page
Short-circuit Clear Zero
5.1
<
SETUP>
page
Press [Setup] key to enter <SETUP>page.
In <SETUP> page, the Instrument does not display measurement result and comparator
result, testing is not in progress.
The <SETUP> page includes following setup
:
Trigger Mode [TRIGGER]
Measurement Range [RANGE]
Measurement Speed [SPEED]
BEEP Feature
Soft-key CLEAR ZERO – Clear Zero Correction
R-COMP – AC Resistance Sorting Methods
V-COMP– DC Voltage Sorting Methods
R-NOMINAL – Input Resistance Nominal Value
R-LOWER – Input Resistance Lower Limit
R-UPPER – Input Resistance Upper Limit
V-NOMINAL – Input Voltage Nominal Value
V-LOWER – Input Voltage Lower Limit
V-UPPER – Input Voltage Upper Limit
Some settings can be also set up in <MEAS DISPLAY> page. Please refer to the chapter:
<Meas> Key
Figure 5‐1 <SETUP>page
5.2
Short
‐
circuit
Clear
Zero
[CLEAR
ZERO]
SCPI Command:
CORR:SHOR
Press [Setup] key to enter < SETUP> page, and then press [CLEAR ZERO] soft-key to
perform short-circuit clear zero.