20
AT4202/4204/4208 User's Guide
Step 3
Use soft keys to select
Soft Key
Function
UPPER
VALUE
Use touch screen to type in
RESET
Restore Defaults
5.1.6
[LOW] setting
n
Steps to set Lowlimit
Step 1
Press [Setup] key to enter < SETUP > page
Step 2
Use cursor keys to select [-200.0] field
Step 3
Use soft keys to select
Soft Key
Function
UPPER
VALUE
Use touch screen to type in
RESET
Restore Defaults
5.1.7
[FAT SET] setting
n
Steps to set FAT
Step 1
Press [Setup] key to enter < SETUP > page
Step 2
Use cursor keys to select [FAT SET] field
Step 3
Use soft keys to select
Soft Key
Function
MICRO
SD
Select MicroSD Memory Card
USB
DISK
Select USB FAT Memory
5.1.8
[INTERVAL
n
Steps to set interval
Step 1
Press [Setup] key to enter < SETUP > page
Step 2
Use cursor keys to select [INTERVAL]
Step 3
Use soft keys to select
Soft Key
Function
UPPER
VALUE
Use touch screen to type in
RESET
Restore Defaults
Note:
First select [ USB DISK ] or [MICRO SD] , and then insert the MICRO SD card or
USB
memory card, the instrument automatically starts recording data.
Example :
Instrument current time is :2013 -11-20 15:30:35
Memory created under the root directory folder name : 20131120 , file name :
153035.CSV