13
Start up
3.11 Setup
Under test state, press Setup key,it will appear the following page:
Press
to select turn page and display setup option.
Press ESC key to return to setup page.
Press Enter key to enter sub menu setup page.
Press Clear key to enter clear zero page, instrument will provide OPEN or SHORT option
according to current state of test terminal:
Press ESC key to return to test state
Perform:
At this moment, please OPEN circuit or SHORT circuit your test clip, instrument will AUTO
check test terminal state,and provide OPEN and SHORT option.
Press Enter key to begin clearing, this process lasts for about 20 seconds, please OPEN circuit
or SHORT circuit your test clip (if previous state is SHORT clear, please OPEN circuit your test
clip; if previous state is OPEN clear, please SHORT circuit your test clip, press Enter again and
wait for about 20 seconds, then clearing finishes and return to test state.)
Press ESC key to terminate clearing process.
After clearing is completed, clearing value is saved in the internal memory of the instrument.
Details
:
When performing clear zero,it possibly appear FAIL (FAIL),this maybe caused by non using of
low resistance short circuit or non reliable contact, please re short circuit and then perform
clear zero.
Instrument will clear all ranges under all frequencies, it will save data in NVM (nonvolatile
memory). Change frequency to test under same test condition, do not reperfrom clear zero.
CLEAR
Current option is glittered
OPEN
1.0000
Current option is glittered
STD