4
Type of Sensing :
Induction coil sensing
for AC current.
Temperature Coefficient :
0.2x(Spec Acc'y)/
℃
, < 18
℃
or > 28
℃
Maximum Output Impedance :
75
Ω
.
Maximum Jaw Opening :
30mm.
Maximum Conductor Size :
29mm diameter.
Size :
72mm (W) x 148mm (L) x 36mm(D)
W/O cable.
Weight :
250grams.
Indoor use.
Maximum Altitude :
2000 meter.
Accessories :
Manual , Carrying Case
.
Precautions and Preparations for
measurement
1. Do not apply the voltage to the output plugs.
2. Do not use or store this instrument in a
high temperature or high humidity
environment and do not store the unit in
direct sunlight.
5
3. Do not measure current before the unit
is not combined with DMM.
4. If the instrument is used near noise gen-
erating equipment , be aware that output
voltage may become unstable or get more
errors.
THIS INSTRUMENT MUST NOT BE
USED ON UNINSULATED CONDUCT-
ORS AT A VOLTAGE GREATER THAN
250V ac/dc.
Electriacl Specification
Accuracy is
±
(%r number of
Ampere) at 23
℃
±
5
℃
, less than 75% R.H.
Range
Measure
Output
Accuracy
400A
3Amp
3mV
30Amp
30mV
350Amp
350mV
400Amp
400mV
±
(3.2% + 1A)
50 ~ 60Hz
±
(1.9% + 0.5A)
50 ~ 60Hz