15
Insulation resistance Measurement
Function
Range
Accuracy *
4.000M
Ω
40.00M
Ω
±(
1.5%+5dgt)
400.0M
Ω
4000M
Ω
±(
3%+5dgt)
4.1G
Ω
~20.0G
Ω
±(
10%+3dgt)
Insulation
Resistance
* Above specifications only apply when high quality silicone
leads with test clips are being used with no hands touch.
Test Voltage vs. Maximum resistance range :
50V/50.0M
Ω
,100V/100.0M
Ω
,250V/250.0M
Ω
,500V/500M
Ω
and
1000V/20.0G
Ω
.
Test Voltage vs. Minimum resistance (with test
current=1mA) :
50V/50k
Ω
,100V/100k
Ω
,250V/250k
Ω
,500V/500k
Ω
and
1000V/1M
Ω
.
Test Voltage Accuracy :
-0%,+20%
Short Circuit Test Current:1mA(nominal)
Auto discharge function :
discharge time<1 sec for C
≦
1
µ
F
Maximum Capacitive load :
Operable with up to 1
µ
F load
Live Circuit Detection :
if
≧
30V ac/dc at inputs, test inhibited
Function
Range
Accuracy
40.00
Ω
±(
1.5%+5dgt)*
400.0
Ω
4000
Ω
40.00k
Ω
±(
1.5%+3dgt)
Earth-bond
Resistance
*<1.00
Ω
add 3dgt
Open Circuit Test Voltage : >4.0V,<8V
Short Circuit Current : >200.0mA
Live Circuit Detection: if
≧
2V ac/dc at inputs, test inhibited.
Earth-bond resistance Measurement