Apogee SL-510 Скачать руководство пользователя страница 16

16 

 

 

TROUBLESHOOTING AND CUSTOMER SUPPORT 

Independent Verification of Functionality 

The radiation detector in Apogee SL-510 and SL-610 pyrgeometers is a self-powered device that outputs a voltage 
signal proportional to the radiation balance between the detector and target surface. A quick and easy check of 
detector functionality can be accomplished using a voltmeter with microvolt (µV) resolution. Connect the positive 
lead of the voltmeter to the white wire from the sensor and the negative lead (or common) to the black wire from 
the sensor. Direct the sensor toward a surface with a temperature significantly different than the detector. The 
voltage signal will be negative if the surface is colder than the detector and positive if the surface is warmer than 
the detector. Placing a piece of tinfoil in front of the sensor should force sensor voltage signal to zero. 

The thermistor inside SL-510 and SL-610 pyrgeometers yields a resistance proportional to temperature. A quick 
and easy check of thermistor functionality can be accomplished with an ohmmeter. Connect the lead wires of the 
ohmmeter to the blue and green wires from the sensor. The resistance should read 30 kΩ at 25 C. If the sensor 
temperature is less than 25 C, the resistance will be higher. If the sensor temperature is greater than 25 C, the 
resistance will be lower. Connect the lead wires of the ohmmeter to the red and green wires from the sensor. The 
resistance should read 24.9 kΩ, and should not vary. Connect the lead wires of the ohmmeter to the red and blue 
wires from the sensor. The resistance should be the sum of the resistances measured across the blue and green 
wires, and red and green wires (e.g., 30 kΩ plus 24.9 kΩ at 25 C). 

Compatible Measurement Devices (Dataloggers/Controllers/Meters) 

Apogee SL-510 and SL-610 pyrgeometers have sensitivities in the microvolt range, approximately 0.11 mV per W 
m

-2

 difference between target and detector. Thus, a compatible measurement device (e.g., datalogger or 

controller) should have resolution of at least 0.11 mV to produce longwave radiation resolution of 1 W m

-2

Measurement of detector temperature from the internal thermistor requires an input excitation voltage, where 
2500 mV is recommended. A compatible measurement device should have the capability to supply the necessary 
voltage. 

An example datalogger program for Campbell Scientific dataloggers can be found on the Apogee webpage at 

http://www.apogeeinstruments.com/content/Thermopile-Pyrgeometer-Unamplified.CR1

. 

Modifying Cable Length 

When the sensor is connected to a measurement device with high input impedance, sensor output signals are not 
changed by shortening the cable or splicing on additional cable in the field. Tests have shown that if the input 
impedance of the measurement device is 10 MΩ or higher, there is negligible effect on the radiometer calibration, 
even after adding up to 50 m of cable. Apogee SL-510 and SL-610 pyrgeometers use shielded, twisted pair cable, 
which minimizes electromagnetic interference. This is particularly important for long lead lengths in 
electromagnetically noisy environments. See Apogee webpage for details on how to extend sensor cable length 
(

http://www.apogeeinstruments.com/how-to-make-a-weatherproof-cable-splice/

). 

Signal Interference 

Due to the small voltage signals from the detector, care should be taken to provide appropriate grounding for the 
sensor and cable shield wire, in order to minimize the influence of electromagnetic interference (EMI). In instances 
where SL-510 and SL-610 series pyrgeometers are being used in close proximity to communications (near an 
antenna or antenna wiring), it may be necessary to alternate the data recording and data transmitting functions 
(i.e., measurements should not be made when data are being transmitted wirelessly). If EMI is suspected, place a 
tinfoil cap over the front of the sensor and monitor the signal voltage from the detector. The signal voltage should 
remain stable at (or very near) zero.

Содержание SL-510

Страница 1: ...NTS INC 721 WEST 1800 NORTH LOGAN UTAH 84321 USA TEL 435 792 4700 FAX 435 787 8268 WEB APOGEEINSTRUMENTS COM Copyright 2020 Apogee Instruments Inc OWNER S MANUAL PYRGEOMETER Models SL 510 and SL 610 R...

Страница 2: ...e of Compliance 3 Introduction 4 Sensor Models 5 Specifications 6 Deployment and Installation 8 Cable Connectors 9 Operation and Measurement 10 Maintenance and Recalibration 15 Troubleshooting and Cus...

Страница 3: ...lectrical and electronic products with respect to the restriction of hazardous substances Please be advised that based on the information available to us from our raw material suppliers the products m...

Страница 4: ...om pyrgeometers are an important input to frost prediction models Pyrgeometers are also integrated into net radiometers for measurement of net radiation at the land surface Apogee Instruments SL 510 a...

Страница 5: ...ing longwave radiation and SL 610 for terrestrial surface measurements downward looking sensor measures outgoing longwave radiation SL 510 Upward looking sensor Sensor model number and serial number a...

Страница 6: ...power requirement at 12 V DC Dimensions 27 5 mm height 23 5 mm diameter Mass 90 g 100 g Cable 5 m of six conductor shielded twisted pair wire TPR jacket high water resistance high UV stability flexib...

Страница 7: ...7...

Страница 8: ...mounting on a cross arm an Apogee Instruments model AM 110 mounting bracket is recommended Sensors must be carefully mounted in order to view the desired target surface sky or ground and avoid includi...

Страница 9: ...atalogger end of the cable If a replacement cable is ever needed please contact Apogee directly to ensure ordering the proper pigtail configuration Alignment When reconnecting the sensor arrows on the...

Страница 10: ...ower source VERY IMPORTANT Apogee changed all wiring colors of our bare lead sensors in March 2018 in conjunction with the release of inline cable connectors on some sensors To ensure proper connectio...

Страница 11: ...emperature are controlled independently At each calibration set point detectors are held at a constant temperature while the blackbody cone is controlled at temperatures below approximately 20 C for f...

Страница 12: ...nt where an excitation voltage is input across the bridge resistor and an output voltage is measured across the thermistor Thermistor wiring for SL 510 Serial Number 1074 and above or with cable conne...

Страница 13: ...4 B 2 21451 x 10 4 and C 1 26233 x 10 7 Longwave Radiation Measurement The detector output from SL 510 and SL 610 pyrgeometers follows the fundamental physics of the Stefan Boltzmann Law where radiat...

Страница 14: ...st deposition The heater is designed to keep the water liquid and frozen off the silicon filter and does not need to be powered in order to make measurements of longwave radiation However if the filte...

Страница 15: ...ts can be removed with a dilute acid e g vinegar Salt deposits cannot be removed with solvents such as alcohol or acetone 3 Dust and dirt deposition on the window usually a larger problem in windy env...

Страница 16: ...fference between target and detector Thus a compatible measurement device e g datalogger or controller should have resolution of at least 0 11 mV to produce longwave radiation resolution of 1 W m 2 Me...

Страница 17: ...ur factory The warranty does not cover equipment that has been damaged due to the following conditions 1 Improper installation or abuse 2 Operation of the instrument outside of its specified operating...

Страница 18: ...manship Apogee Instruments will repair or replace the items free of charge If it is determined that your product is not covered under warranty you will be informed and given an estimated repair replac...

Отзывы: