MAINTENANCE AND RECALIBRATION 

Blocking of the optical path between the target and detector, often due to moisture or debris on the silicon filter, 
is a common cause of inaccurate measurements. SL-510 pyrgeometers have a domed housing for improved self-
cleaning from rainfall, but the flat filter surface of SL-510 and SL-610 pyrgeometers can become partially blocked in 
three ways: 

1.

 

Dew or frost formation on the silicon filter (can be minimized by powering heater). 

2.

 

Salt deposit accumulation on the filter, due to evaporating irrigation water or sea spray. This leaves a thin 
white film on the filter surface. Salt deposits can be removed with a dilute acid (e.g., vinegar). Salt 
deposits cannot be removed with solvents such as alcohol or acetone.
 

3.

 

Dust and dirt deposition on the window (usually a larger problem in windy environments). Dust and dirt 
are best removed with deionized water, rubbing alcohol, or in extreme cases, acetone. 

Clean filter with a cotton swab dipped in the appropriate solvent. Never use an abrasive material on the silicon 
filter.
 Use only gentle pressure when cleaning the filter with a cotton swab, to avoid scratching the outer surface. 
The solvent should be allowed to do the cleaning, not mechanical force. 

It is recommended that pyrgeometers be recalibrated every two years. See the Apogee webpage for details 
regarding return of sensors for recalibration (

http://www.apogeeinstruments.com/tech-support-recalibration-

repairs/

)

.

Содержание SL-510

Страница 1: ...NTS INC 721 WEST 1800 NORTH LOGAN UTAH 84321 USA TEL 435 792 4700 FAX 435 787 8268 WEB APOGEEINSTRUMENTS COM Copyright 2020 Apogee Instruments Inc OWNER S MANUAL PYRGEOMETER Models SL 510 and SL 610 R...

Страница 2: ...e of Compliance 3 Introduction 4 Sensor Models 5 Specifications 6 Deployment and Installation 8 Cable Connectors 9 Operation and Measurement 10 Maintenance and Recalibration 15 Troubleshooting and Cus...

Страница 3: ...lectrical and electronic products with respect to the restriction of hazardous substances Please be advised that based on the information available to us from our raw material suppliers the products m...

Страница 4: ...om pyrgeometers are an important input to frost prediction models Pyrgeometers are also integrated into net radiometers for measurement of net radiation at the land surface Apogee Instruments SL 510 a...

Страница 5: ...ing longwave radiation and SL 610 for terrestrial surface measurements downward looking sensor measures outgoing longwave radiation SL 510 Upward looking sensor Sensor model number and serial number a...

Страница 6: ...power requirement at 12 V DC Dimensions 27 5 mm height 23 5 mm diameter Mass 90 g 100 g Cable 5 m of six conductor shielded twisted pair wire TPR jacket high water resistance high UV stability flexib...

Страница 7: ...7...

Страница 8: ...mounting on a cross arm an Apogee Instruments model AM 110 mounting bracket is recommended Sensors must be carefully mounted in order to view the desired target surface sky or ground and avoid includi...

Страница 9: ...atalogger end of the cable If a replacement cable is ever needed please contact Apogee directly to ensure ordering the proper pigtail configuration Alignment When reconnecting the sensor arrows on the...

Страница 10: ...ower source VERY IMPORTANT Apogee changed all wiring colors of our bare lead sensors in March 2018 in conjunction with the release of inline cable connectors on some sensors To ensure proper connectio...

Страница 11: ...emperature are controlled independently At each calibration set point detectors are held at a constant temperature while the blackbody cone is controlled at temperatures below approximately 20 C for f...

Страница 12: ...nt where an excitation voltage is input across the bridge resistor and an output voltage is measured across the thermistor Thermistor wiring for SL 510 Serial Number 1074 and above or with cable conne...

Страница 13: ...4 B 2 21451 x 10 4 and C 1 26233 x 10 7 Longwave Radiation Measurement The detector output from SL 510 and SL 610 pyrgeometers follows the fundamental physics of the Stefan Boltzmann Law where radiat...

Страница 14: ...st deposition The heater is designed to keep the water liquid and frozen off the silicon filter and does not need to be powered in order to make measurements of longwave radiation However if the filte...

Страница 15: ...ts can be removed with a dilute acid e g vinegar Salt deposits cannot be removed with solvents such as alcohol or acetone 3 Dust and dirt deposition on the window usually a larger problem in windy env...

Страница 16: ...fference between target and detector Thus a compatible measurement device e g datalogger or controller should have resolution of at least 0 11 mV to produce longwave radiation resolution of 1 W m 2 Me...

Страница 17: ...ur factory The warranty does not cover equipment that has been damaged due to the following conditions 1 Improper installation or abuse 2 Operation of the instrument outside of its specified operating...

Страница 18: ...manship Apogee Instruments will repair or replace the items free of charge If it is determined that your product is not covered under warranty you will be informed and given an estimated repair replac...

Отзывы: