background image

15 

 

 

TROUBLESHOOTING AND CUSTOMER SUPPORT 

Verify Functionality 

Pressing the power button should activate the LCD and provide a real-time PPFD reading. Direct the sensor head toward a 
light source and verify the PPFD reading responds. Increase and decrease the distance from the sensor to the light source to 
verify that the reading changes proportionally (decreasing PPFD with increasing distance and increasing PPFD with 
decreasing distance). Blocking all radiation from the sensor should force the PPFD reading to zero. 

Battery Life 

When the meter is maintained properly the coin cell battery (CR2320) should last for many months, even after continuous 
use. The low battery indicator will appear in the upper left hand corner of the LCD display when the battery voltage drops 
below 2.8 V DC. The meter will still function correctly for some time, but once the battery is drained the pushbuttons will no 
longer respond and any logged measurements will be lost. 

Pressing the power button to turn off the meter will actually put it in sleep mode, where there is still a slight amount of 
current draw. This is necessary to maintain the logged measurements in memory. Therefore, it is recommended to remove 
the battery when storing the meter for many months at a time, in order to preserve battery life.  

Low-Battery Error after Battery Replacement 

A master reset will usually correct this error, please see the master reset section for details and cautions. If a master reset does 
not remove the low battery indicator, please double check that the voltage of your new batter is above 2.8 V, this is the 
threshold for the indicator to turn on. 

Master Reset 

If a meter ever becomes non-responsive or experiences anomalies, such as a low battery indicator even after replacing the 
old battery, a master reset can be performed that may correct the problem. Note that a master reset will erase all logged 
measurements from memory.  

First press the power button so that the LCD display is activated. While still powered, slide the battery out of the holder, 
which will cause the LCD display to fade out. After a few seconds, slide the battery back into the holder. The LCD display will 
flash all segments and then show a revision number (e.g. “R1.0”). This indicates the master reset was performed and the 
display should return to normal. 

Error Codes and Fixes 

Error codes will appear in place of the real-time reading on the LCD display and will continue to flash until the problem is 
corrected. Contact Apogee if the following fixes do not rectify the problem. 

Err 1:

 battery voltage out of range. 

Fix:

 replace CR2320 battery and perform master reset. 

Err 2:

 sensor voltage out of range. 

Fix: 

perform master reset. 

Err 3:

 not calibrated. 

Fix:

 perform master reset. 

Err 4:

 CPU voltage below minimum. 

Fix:

 replace CR2320 battery and perform master reset. 

Modifying Cable Length 

Although it is possible to splice additional cable to the separate sensor of the appropriate MQ model, note that the cable 
wires are soldered directly into the circuit board of the meter. Care should be taken to remove the back panel of the meter in 
order to access the board and splice on the additional cable, otherwise two splices would need to be made between the 

Содержание MQ-100

Страница 1: ...ENTS INC 721 WEST 1800 NORTH LOGAN UTAH 84321 USA TEL 435 792 4700 FAX 435 787 8268 WEB APOGEEINSTRUMENTS COM Copyright 2016 Apogee Instruments Inc OWNER S MANUAL QUANTUM METER Models MQ 100 MQ 200 and MQ 300 Series ...

Страница 2: ... Certificate of Compliance 3 Introduction 4 Sensor Models 5 Specifications 6 Deployment and Installation 9 Operation and Measurement 10 Maintenance and Recalibration 14 Troubleshooting and Customer Support 15 Return and Warranty Policy 17 ...

Страница 3: ...nt control and laboratory use EMC requirements EN 50581 2012 Technical documentation for the assessment of electrical and electronic products with respect to the restriction of hazardous substances Please be advised that based on the information available to us from our raw material suppliers the products manufactured by us do not contain as intentional additives any of the restricted materials in...

Страница 4: ...ized nature of radiation A quantum refers to the minimum quantity of radiation one photon involved in physical interactions e g absorption by photosynthetic pigments In other words one photon is a single quantum of radiation Typical applications of quantum sensors include incoming PPFD measurement over plant canopies in outdoor environments or in greenhouses and growth chambers and reflected or un...

Страница 5: ...easurements All sensors along the length of the line are connected in parallel and as a result Apogee line quantum meters display PPFD values that are averaged from the location of the individual sensors Sensor model number and serial number are located on a label on the backside of the handheld meter MQ 100 Meter w integral sensor MQ 200 Meter w separate sensor MQ 300 Series Meter w separate line...

Страница 6: ...eatability Less than 1 Long term Drift Non stability Less than 2 per year Non linearity Less than 1 up to 3000 µmol m 2 s 1 Response Time Less than 1 ms Field of View 180º Spectral Range 410 to 655 nm wavelengths where response is greater than 50 of maximum see Spectral Response below Directional Cosine Response 5 at 75º zenith angle see Cosine Response below Temperature Response 0 06 0 06 per C s...

Страница 7: ...sors were compared to the spectroradiometer The spectroradiometer was mounted external to the temperature control chamber and remained at room temperature during the experiment Mean spectral response of six SQ series quantum sensors error bars represent two standard deviations above and below mean compared to PPFD weighting function Spectral response measurements were made at 10 nm increments acro...

Страница 8: ...our reference thermopile pyranometers with solar zenith angle dependent factors applied to convert total shortwave radiation to PPFD Blue points represent the AM response and red points represent the PM response Directional or cosine response is defined as the measurement error at a specific angle of radiation incidence Error for Apogee SQ series quantum sensors is approximately 2 and 5 at solar z...

Страница 9: ... with a different option for mounting the sensor to a horizontal plane MQ 300 series line quantum sensors are leveled using the built in bubble level located in the handle of the sensor In addition to leveling all sensors should also be mounted such that obstructions e g weather station tripod tower or other instrumentation do not shade the sensor The AL 210 leveling plate is recommended for use w...

Страница 10: ...on is also used to view and scroll through the logged measurements on the LCD display Calibration To choose between sunlight and electric light calibration push the mode button once and use the up down buttons to make the appropriate selection SUN or ELEC Once the desired mode is blinking press the mode button three more times to exit the menu Logging To choose between manual or automatic logging ...

Страница 11: ... T5 cool white fluorescent lamps will read approximately 12 low in sunlight In addition to PPFD measurements Apogee MQ series quantum meters can also be used to measure yield photon flux density YPF photon flux weighted according to the plant photosynthetic action spectrum McCree 1972 and summed YPFD is also expressed in units of µmol m 2 s 1 and is similar to PPFD but is typically more closely co...

Страница 12: ...al Halide 6 0 6 4 High Pressure Sodium 0 8 7 2 Blue LED 448 nm peak 20 nm full width half maximum 12 7 8 0 Green LED 524 nm peak 30 nm full width half maximum 8 0 26 2 Red LED 635 nm peak 20 nm full width half maximum 4 8 6 2 Red Blue LED Mixture 85 Red 15 Blue 2 4 4 4 Red Green Blue LED Mixture 72 Red 16 Green 12 Blue 3 4 0 2 Cool White Fluorescent LED 4 6 0 6 Neutral White Fluorescent LED 6 7 5 ...

Страница 13: ...or this effect underwater measurements are only relative which makes it difficult to compare light in different environments The original quantum sensors covered in this manual have an immersion effect correction factor of 1 08 If you wish to use your meter to make measurements underwater simply multiply the measured number by the immersion effect 1 08 When making underwater measurements only the ...

Страница 14: ... the sky must be completely clear as reflected radiation from clouds causes incoming radiation to increase above the value predicted by the clear sky calculator Measured values of PPFD can exceed values predicted by the Clear Sky Calculator due to reflection from the sides and edges of clouds This reflection increases the incoming radiation The influence of high clouds typically shows up as spikes...

Страница 15: ...icator please double check that the voltage of your new batter is above 2 8 V this is the threshold for the indicator to turn on Master Reset If a meter ever becomes non responsive or experiences anomalies such as a low battery indicator even after replacing the old battery a master reset can be performed that may correct the problem Note that a master reset will erase all logged measurements from...

Страница 16: ...ertain applications It is possible to convert the PPFD value from a quantum sensor to other units but it requires spectral output of the radiation source of interest Conversion factors for common radiation sources can be found on the Unit Conversions page in the Support Center on the Apogee website http www apogeeinstruments com unit conversions A spreadsheet to convert PPFD to energy flux density...

Страница 17: ...Improper installation or abuse 2 Operation of the instrument outside of its specified operating range 3 Natural occurrences such as lightning fire etc 4 Unauthorized modification 5 Improper or unauthorized repair Please note that nominal accuracy drift is normal over time Routine recalibration of sensors meters is considered part of proper maintenance and is not covered under warranty Who is Cover...

Страница 18: ...nsible for any direct indirect incidental or consequential damages including but not limited to loss of income loss of revenue loss of profit loss of wages loss of time loss of sales accruement of debts or expenses injury to personal property or injury to any person or any other type of damage or loss This limited warranty and any disputes arising out of or in connection with this limited warranty...

Отзывы: