Operation
Setup procedure
3DLevelScanner Operating Instruction. © 2009 APM all rights reserved
23
6.5.6
Performing Full Calibration adjustment
P
roceed as follows
:
full calibr.
006
1.75m
distance flange
to max. level
This function is used to enter the distance from the bottom of flange (measurement
reference point) to the maximum level (100%).
CAUTION!
The maximum level may not project into the blocking distance (BD). If the
BD is compromised, it may cause device malfunction. After basic calibration,
enter a safety distance (SD) in the safety distance (015) function. If the level is within
this SD, the 3DLevelScanner M signals a warning or an alarm, depending on your
selection in the 'in safety distance' (016) function.
1.
E
nter the appropriate distance value in m and save with
[E]
.
2.
S
ave the settings with
[E]
At this point, the basic adjustment is complete.