Performance

 

Operating Range 

0 - 17 ft. / 0 - 5 m

 

 

0 - 50 ft. / 0 - 15 m

 

 

0 - 100 ft. / 0 - 30 m 

 

 

Transducer Frequency   

 

 

PRL: 6.3 MHz

 

 

 

 

 

 

 

PRS: 6.3 MHz & 26 MHz

 

Target Dielectric Constant 

 

 

ε

r

 > 2

 

Beam Spread   

 

 

 

Teflon Antenna: 6° from center

 

 

 

 

 

 

 

316L SS Horn: 3° from center

Accuracy

 

Accuracy 

 

 

 

 

±0.25% of measured range

 

Resolution 

 

 

 

 

6.1 μA

Environmental

 

Housing Operating Temperature 

 

-40 to 140°F   (-40 to 60°C)

 

Process/Probe Temperature

 

 

Standard Temperature  

 

-40 to 176°F   (-40 to 80°C)

 

 

High Temperature 

 

 

-40 to 350°F   (-40 to 177°C)

 

 

Sanitary Temperature   

 

-40 to 400°F   (-40 to 204°C)

 

 

Extreme Temperature   

 

-40 to 572°F   (-40 to 300°C)

 

Maximum Process Pressures

 

 

Standard Temp/Pressure 

 

5 bar (72.5 psi)

 

 

Standard Temp/High Pressure  

70 bar (1015.3 psi)

 

 

High/Sanitary/Extreme Temps   

2 bar (29 psi)

 

Housing Enclosure Rating 

 

 

NEMA 6 / IP 68

Electrical

 

Supply Voltage  

 

 

 

12-30 VDC; 20-35 VDC (HART)

 

 

 

 

 

 

 

115 VAC @ 60 Hz / 230 VAC @ 50 Hz

 

Current Draw   

 

 

 

70 mA max, 1.68 W @ 24 VDC

 

 

 

 

 

 

 

1.7 VA for 115/230 VAC

 

Materials of Construction

 

Housing 

 

 

 

 

Aluminum, 316L Stainless Steel

 

Antenna 

 

 

 

 

Teflon, 316L Stainless Steel Horn

Mounting

 

Teflon Antenna  

 

 

 

1.5”  or 2” NPT

 

Sanitary Teflon Antenna 

 

 

1.5” or 2” Sanitary Tri-Clamp

 

316L SS 6” Ø Horn 

 

 

 

3” NPT

 

316L SS 13” Ø Horn 

 

 

 

13” Ø Flange

Communication

 

Output  

 

 

 

 

4-20 mA (shared common or isolated); HART

 

Calibration & Diagnostics 

 

 

RS-232, RS-485 Modbus, HART

• 

Specifications

Содержание True Echo PRL Series

Страница 1: ...APG R Doc 9002986 Part 200213 Rev B 01 2019 True EchoTM Pulse Radar Level Transmitter User Manual Series PRL for Liquids and Series PRS for Solids...

Страница 2: ...Specifications 2 Model Number Configurators 3 4 Electrical Terminals and System Wiring Diagrams 5 6 Chapter 2 Installation and Removal Procedures and Notes 7 Tools Needed 7 Installation Notes 7 Mounti...

Страница 3: ...ly programmable via RS 232 RS 485 Modbus or Hart communications Reading your label Every APG instrument comes with a label that includes the instrument s model number part number serial number and a w...

Страница 4: ...material and workmanship under normal use and service of the product for 24 months For a full explanation of our Warranty please visit https www apgsensors com about us terms conditions Contact Techni...

Страница 5: ...2 125 53 97 7 30 185 42 4 00 101 6 10 30 261 62 18 60 472 44 1 00 25 40 3 ANSI FF Flange Not Provided Radar with Horn Antenna Radar with Sanitary Antenna Radar with Standard Antenna Cover Tool Access...

Страница 6: ...40 to 400 F 40 to 204 C Extreme Temperature 40 to 572 F 40 to 300 C Maximum Process Pressures Standard Temp Pressure 5 bar 72 5 psi Standard Temp High Pressure 70 bar 1015 3 psi High Sanitary Extreme...

Страница 7: ...nsion LT Teflon Rod with 1 5 in built in extension ST Sanitary Tri Clamp mount Teflon Ant HT High Temperature Radar Tefon Ant PT High Pressure Tefon Rod Ant 70 bar S6 316L Stainless Steel 6 Horn HS Hi...

Страница 8: ...wire Isolated 4 20 mA V024 12 30 VDC 3 wire 4 20 mA w shared common V20L 20 35 VDC 2 wire HART C PC Communications C2 RS 232 C4 RS 485 CH HART Note CH HART PC Communications requires V20L 20 35 VDC S...

Страница 9: ...1 5 4 3 RS 232 485 VAC L1 VAC L2 N Ground 5 115 230 VAC Power Input Card 4 20mA L R True Echo Terminals B RX A TX Network Node Communication Out Ground A TX B RX Supply Voltage 115 230 VAC 1 Phase Pa...

Страница 10: ...inals Red 1 2 3 4 5 6 7 8 9 WHT or BLK SHIELD L1 8 L2 N 7 6 5 4 3 A TX 2 B RX 1 PR S Terminals To PC USB Port Red WHT or BLK L1 8 L2 N 7 6 5 4 3 A TX 2 B RX 1 RS485 to USB Converter Shield Communicati...

Страница 11: ...pipe please see our guide to stand pipes on our website http www apgsensors com about us blog how to install a stand pipe You should not need any tools to install your True EchoTM series Radar Level...

Страница 12: ...VAC or VDC control power and communication power to the radar is off Unscrew top lid and detach wires from terminals Disconnect conduit from radar Remove the radar and store it in a dry place at a tem...

Страница 13: ...ectric Materials mode is OFF To toggle the Low Dielectric Materials mode press and hold the Calibration Switch button until the Status LED cycles from Yellow to Red and then off see Terminals on page...

Страница 14: ...with instructions Phone 888 525 7300 Email sales apgsensors com Online chat at www apgsensors com Please have your radar s part number and serial number available See Warranty and Warranty Restriction...

Страница 15: ...11 Tel 1 888 525 7300 Fax 1 435 753 7490 www apgsensors com sales apgsensors com...

Страница 16: ...Automation Products Group Inc Tel 1 888 525 7300 Fax 1 435 753 7490 www apgsensors com sales apgsensors com APG R...

Отзывы: