21
USER'S GUIDE
Chapter 3.
Glucose Testing
Step 1:
After you have performed the previous section, "
Coding the Meter
," the meter
will now self-test the environment temperature. If the temperature is below or
above operation range, the screen will display the flashing symbol of thermometer
"
". When the temperature is within the testing range of 18°C to 38°C
(64°F to 100°F), the screen will show a symbol "
"and the meter is now ready
for testing.
IMPORTANT:
1. Each time the GlucoSure
®
Plus Blood Glucose Meter is used, the current code
will appear on the display screen when turned on. Verify that the code displayed
on screen matches the code number printed on the test strip package before
testing. The code needs to be set only once for each package of test strips. The
meter will memorize the code until it is changed. Code number displayed on
screen not matching the code number printed on the test strip package will
create an inaccurate test result.
2. If the screen shows "Code----" (meaning that the Meter is not coded), or if the
code number on the screen does not match the code number on the test strip
package, refer to Chapter 3, p.20 "
Coding the Meter
"
.