: Values determined according to noise test code given in ISO 80601-2-70:2015, using the
basic standards ISO 3744 and ISO 7779.
7.4 Pressure Accuracy
Maximum static pressure variation at 10 cmH
2
O (10 hPa) according to ISO 80601-2-70:2015
15mm Light grey Tubing
W
Without Humidification
±1.0 cmH
2
O (±1.0 hPa)
With Humidification
±1.0 cmH
2
O (±1.0 hPa)
Static pressure accuracy has a measurement uncertainty of ± 0.15 cm H
2
O
Maximum dynamic pressure variation according to ISO 80601-2-70:2015
Device without/with Humidification and 15mm Light grey Tubing
Pressure(cmH2O)
10 BPM
15 BPM
20 BPM
< 10 cmH
2
O
1.5 / 2.0
2.5 / 3.0
3.0 / 3.5
≥ 10.0 to 20 cm H O
2.0 / 2.5
3.5 / 4.0
4.0 / 4.8
Dynamic pressure accuracy has a measurement uncertainty of ± 1.03 cm H
2
O
Results are expressed as STPD (Standard Temperature and Pressure, Dry).
7.5 Maximum Flow Rate
Pressure (cm H
2
O)
Test pressures (cm H
2
O )
4
8
12
16
20
15 mm
tubing
Flow (LPM)
>30
>30
>30
>30
>30
Results are expressed as STPD (Standard Temperature and Pressure, Dry).
7.6 Maximum Single Fault pressure
When assessing the relative risks of using this equipment, the Physician should understand that this
device can deliver pressures up to 20 cmH
2
O. In the event of certain fault conditions, a maximum
pressure is 40 cmH
2
O. Results are expressed as STPD (Standard Temperature and Pressure, Dry).