4. OPERATIONS
NOTE:
Always read the operating instructions before use.
4.1 GENERAL OPERATION
1. Switch on the main power switch found on the side
of the pump.
2. Once the pump begins delivering air into the
mattress it will take approximately 30 minutes to
fully inflate the mattress.
3. Every time when mattress is first setup for use, It
will be forced to execute Maxfirm for the quickest
inflation. The low-pressure indicator (yellow LED)
will light up when the mattress is not fully inflated at
initial inflation. If the appropriate pressure is
reached, the low-pressure indicator (yellow LED)
will go off.
4. When the initial inflation (Maxfirm process) is
completed, the sy
stem will enter the “alternating
mode
”.
5. According to the weight and height of the patient,
adjust the pressure setting to the most comfortable
level without bottoming out, then the pressure in
mattress will slowly increase to the intended value
after the air mattress is ready to use.
NOTE:
Every time when mattress is first
setup for use, It will be forced to
execute Maxfirm for the quickest
inflation. User can then adjust air
mattress to the desired softness
afterward.
4.2 EMERGENCY CPR OPERATIONS
CPR must be performed on a firm surface. Therefore, if an
emergency CPR situation occurs with the patient on the
mattress, the mattress must be quickly deflated. This is done
by quickly pulling on the CPR tag located at the head of the
mattress on the patient
’s right-side. The quick connector found
from the pump unit can also be disconnected to hasten deflation.