LabSen762 Blade Spear pH Electrode

 

User Manual

 

 

LabSen electrochemical sensors are premium pH electrode with manufacturing technology and key 
components  imported  from  Switzerland.  LabSen762  places  the  tapered  glass  membrane  in  a 
stainless  steel  shaft  with  blade.  This  probe  is  suitable  for  pH  measuring  of  fish,  meat  and  meat 
products. 

 

This Electrode has following features: 

 

 Adopts food grade stainless steel blade and sheath, offers strong protection, comply with food 

hygiene requirements, suitable for food pH testing.

 

 Blue gel inner solution, does not flow and will not cause bubble.

 

 Maintenance-free polymer electrolyte, non-diaphragm structure.

 

 Long life reference system, has better stability and service life.

 

 

Technical Data

 

Measuring Range

 

(0 ~ 14) pH

 

Electrolyte

 

Polymer

 

Temperature Range

 

(0 ~ 80)°C

 

Soaking Solution

 

3M KCl

 

Membrane Shape

 

Tapered

 

Electrode Dimension

 

(Ø12×115) mm

 

Reference

 

Long Life

 

Connector

 

S7

 

Junction

 

Single Pore

 

Cable

 

S7-BNC / Ø3×1m

 

 

Usage and Maintenance

 

2.1 

Connect electrode to S7-BNC cable, and then connect BNC plug to the input on the pH meter.

 

2.2 

When  measuring,  please  unscrew  the  bottle  cap,  pull  out  the  electrode  and  rinse  it  with 
deionized water. After using, please put the electrode back into the bottle and screw tight the 
cap.

 

 
 
 
 
 
 
 
 
 
 

2.3 

Please forcibly even when piercing into meat (especially frozen meat); do not bend it to prevent 
damage of the tapered membrane. Wash the electrode and the stainless steel sheath with soapy 
water after measurement, the stainless blade can be screwed off and cleaned if necessary.

 

2.4 

The  electrode  must  be  soaked  in  Protelyte  solution  for  storage.  Because  after  long  time 
exposure in air, Polymer electrolyte will cause shrinkage.

 

2.5 

The electrode’s measuring tip should be soaked in the soaking bottle containing certain amount 
of Protelyte soaking solution to keep the membrane hydrated and junction unblocked. Clean the 

Before pulling out or putting back the electrode, 
make sure to fully loosen the blue cap on the 
storage sleeve so that the electrode can move 
in and out smoothly.

 