TECHNICAL SPECIFICATIONS 

 

2.1 

Technical Parameters 

Measuring Method 

 ISO 7027 and DIN EN 27027 compliant nephelometric method

90°

 

Light Source 

Infrared light emitting diode (860±30nm) 

Measuring Range 

0 ~ 1000 NTU, automatic range selection: 

0.01~19.99 NTU 

20.0~99.9 NTU 
100~1000 NTU 

Accuracy 

≤ ±2 % of r stray light 

Repeatability  

≤ ± 1% of reading or 0.02 NTU (FTU), the greater of the two 

Resolution 

0.01 / 0.1 / 1 NTU(FNU) 

Stray Light 

≤0.02 NTU(FNU) 

Calibration Standards 

AMCO Polymer or Formazin Solution:   

0, 20, 100, 400 and 800 NTU (FNU) 

Detector 

Silicon photovoltaic 

Measurement Mode 

Normal measurement and average measurement mode 

Display 

TFT color screen 

Sample Vials 

Φ25×60 mm, high borosilicate glass with lid 

Sample Volume 

18 ml 

Power Supply 

4× “AA” Alkaline Batteries 

Operating Condition 

Temperature: 0 to 50°C

32°F to 122°F

Relative humidity

   

0 to 90% at 30°C, 0 to 80% at 40°C, 0 to 70% at 50°C, no condensation 

Storage Condition 

Instrument: -40~60

  (-40~140

Calibration Solution: 5~30

  (41~86

Instrument sealing grade 

IP67    

Certificates 

  CE 

Limited Warranty 

2 years 

Dimension and Weight 

  Meter: (90×203×80) mm/385g 

Test Kit: (310×295×110) mm/1.5kg 

     

  

 

 

Содержание TN400

Страница 1: ...1 TN400 Portable Turbidity Meter Instruction Manual APERA INSTRUMENTS LLC aperainst com...

Страница 2: ...Display Mode 6 3 4 Keypad 7 3 5 Power Supply 7 4 Meter Calibration 8 4 1 Preparation for Calibration 8 4 2 Calibration Procedure Take 0 NTU and 20 NTU as an example 9 4 3 Notes for Calibration 10 5 T...

Страница 3: ...characteristics Smart functions such as automatic calibration operating instructions automatic power off low voltage display etc Large TFT color screen with blue background in measuring mode green bac...

Страница 4: ...s AMCO Polymer or Formazin Solution 0 20 100 400 and 800 NTU FNU Detector Silicon photovoltaic Measurement Mode Normal measurement and average measurement mode Display TFT color screen Sample Vials 25...

Страница 5: ...vial holder Close the cover when measuring Dust proof plug Take off the plug when measuring Housing Sample vial holder Display Calibration vials or sample vials Keypad Positioning mark Align the mark...

Страница 6: ...0 NTU Carrying case TN400 Turbidimeter Microfiber cloth Screwdriver to remove battery cover Silicone oil 10 ml AA Alkaline battery 4 Sample vial 6 Instruction manual Diagram 2 Measurement mode Measure...

Страница 7: ...erform single shot measurement long press the key to perform continuous measurement In calibration mode press the key to confirm the calibration In calibration mode press the key to select standard so...

Страница 8: ...your supplier to replace See the detail in Section 7 Replacement Parts b Replace zero turbidity solution Replace with 0 0 NTU calibration solution open the vial cap pour out the original solution add...

Страница 9: ...the vial to make the solution even For polymer calibration solutions that have not been used for a long time slowly flip the vial twice and let it stand for 2 minutes Be careful not to shake the solu...

Страница 10: ...00 NTU calibration vial in the sample vial holder and press to calibrate After calibration is done press to confirm repeat the steps to calibrate 400 NTU and 800 NTU After calibration is done press to...

Страница 11: ...ation solution If the wrong solution is selected for calibration the measurement will be completely wrong To fix this problem simply re calibrate the meter selecting the correct calibration solutions...

Страница 12: ...s of continuous measurement displaying 1 data every 8 seconds and will display the average value at the end see diagram 6 The average measurement mode can be used for observing the stabilizing process...

Страница 13: ...d for measurement Please do not wash the sample vial holder as it may damage the optical structure 6 WARRANTY We warrant this instrument to be free from defects in material and workmanship and agree t...

Страница 14: ...L 1 bottle T200 1 20 100 400 800 NTU AMCO polymer Compatible with TN400 TN480 4 bottles T500 3 Sample vials with lid 25 60mm 1 6 6 pcs TN400 S3 Silicone oil 10mL 1 bottle APERA INSTRUMENTS LLC Address...

Отзывы: