6.  Notes 

6.1 The tester adopts 1413 µS and 12.88 mS standard calibration solutions. Users can 

use 1-2 point calibrations as needed. For most circumstances, calibrating in 1413 µS to 

complete 1

st

 point calibration will meet testing requirements. 

6.2 The tester has self-diagnosis functions: 

Symbol 

Self-Diagnosis information 

How to fix 

 

 

Wrong calibration solution, which 
exceeds the recognizable range of 
the meter. 

1. Check if calibration solution is 
correct 
2. Check if probe is damaged. 

 

 

 is pushed before measurement 

is stable (

 comes up) 

Wait for the smile icon to stay, 

and then short press  

    

 

6.3 The tester has already been calibrated after manufacture. Usually, users can use 

the tester right away, or test it in the calibration solutions to test its accuracy. When 

error is large, calibrate it before using. 

6.4 We recommend replacing new calibration solutions after 5-10 times of calibrations 

to keep the solution’s accuracy. 

 

7.  Parameter Setting 

7.1 Parameter setting reference chart: 

Symbol 

Parameter Setting Items 

Code 

Factory Default 

P1 

TDS Factor 

0.40 to 1.00 

0.71 

P2 

Select Temperature Unit 

°C - °F 

°C 

P3 

Restore to factory default 

No 

– Yes 

No 

 

7.2 How to setup parameters: 

When turned off, long press 

  to enter setup 

 short press  

 to switch  

P1-P2-P3

 Short press 

, parameter flickering

short press 

 to choose, 

short press 

 to confirm parameter selection

Long press 

 to switch off. 

 