3.  Complete Kit 

 

 

4.  Calibration 

4.1 Rinse the probe in distilled water and dry it. Short press 

 to 

turn on the tester. 

4.2 Long press 

 to enter calibration mode; Short press 

 to exit. 

4.3 Dip the probe into 1413 

µ

S calibration solution. Stir gently, leave it 

to stand, 

LCD displays 

(Diagram 3), short press 

 to complete 1-point calibration. 

The tester returns to measuring mode, and calibration icon 

 appears at the 

button left of LCD. 

4.4 Rinse probe in distilled water and dry it. Follow the steps in 4.2 - 4.3 to 

complete 2

nd

 point calibration in 12.88 mS calibration solution, tester returns to 

measuring mode, calibration icons 

 

 display on bottom left of LCD. 

 

5.  TDS Measurement 

5.1 Short press  

 to turn on tester. Rinse probe in distilled water and dry it. 

5.2 Stir probe in the sample solution gently, leave it to stand. Get readings after the 

smile icon comes up and stays. 

 

Diagram - 2 

 

Diagram - 3 

 