1. What’s in the Kit

2. Keypad Functions

Short press:   

2 seconds         Long press:   

2 seconds

1. Short press to turn on the tester and long press to turn 

off the tester.

2. When turned off, long press to enter parameter setting. 
3. In measurement mode, short press to turn on backlight.
4. In calibration mode, short press to cancel calibration.

1. In measurement mode, short press to switch between 

pH

 and 

mV

.

2. In  settings,  short  press  to  change  parameters 

(Unidirectional).

1. Long press to enter calibration mode.
2. In calibration mode, short press to confirm calibration.
3. When reading is locked, short press to unlock.

Battery Cap

LCD

Buttons

Probe ring

Probe Cap

Lanyard

Calibration Vials

pH4.00

pH buffer

Vials holder

3M KCL Storage Solution

PH60S Tester

pH7.00

pH buffer

pH10.01

pH buffer

Swiss spear 

pH probe

Содержание PH60S

Страница 1: ...1 Premium Series PH60S Spear pH Tester pH mV Temperature User Manual IP67 ISO 9001 2015 APERA INSTRUMENTS LLC aperainst com...

Страница 2: ...aning 6 7 Probe Storage 6 8 Parameter Settings 6 9 ORP Measurement 7 10 Technical Specifications 8 11 Other Specifications 8 12 Probe Replacement 9 13 Notes 9 14 Battery Replacement 10 15 Troubleshoot...

Страница 3: ...hort press to cancel calibration 1 In measurement mode short press to switch between pH and mV 2 In settings short press to change parameters Unidirectional 1 Long press to enter calibration mode 2 In...

Страница 4: ...pH 7 00 buffer solution first and make a quick stir in the solution then hold still 4 1 4 Long press to enter calibration mode screen will turn green Short press if you decide to quit calibration and...

Страница 5: ...quirement 0 1 pH 2 pt Option A 1st pt 7 00 pH 2nd pt 4 00 pH or 1 68 pH Option A 1st pt 6 86 pH 2nd pt 4 01 pH or 1 68 pH Range 7 00 pH Option B 1st pt 7 00 pH 2nd pt 10 01 pH or 12 45 pH Option B 1st...

Страница 6: ...y could damage the pH probe Pure water is only for rinsing the probe 5 5 Pure Water pH Measurement When testing pure water like tap water drinking water RO water or distilled water it will take longer...

Страница 7: ...on screen screen turns red 8 3 3 Automatic Lock P4 Select On to activate auto lock function When reading is stable for more than 10 seconds the tester will lock the value automatically and HOLD icon w...

Страница 8: ...mp Measuring Range 0 50 C 32 122 F Resolution 0 1 C Accuracy 0 5 C 11 Other Specifications Screen 3 color LCD screen Blue Measurement Green Calibration Red Alarm Reading Lock HOLD Low Voltage Warning...

Страница 9: ...PH60S E spear pH probe for soft solid samples AI1203 PH60F E flat pH probe for surface test AI1201 PH60 E regular pH probe for general water solutions AI3711 E PH60 DE double junction pH probe for gen...

Страница 10: ...E Battery MUST FACE UP WRONG INSTALLATION OF BATTERIES WILL CAUSE DAMAGE TO THE TESTER AND POTENTIAL HAZARDS Loosen the battery cap lock Pull off the battery cap Slide and unlock battery compartment O...

Страница 11: ...t s within the 6 month electrode warranty contact us for warranty fulfillment If there is visible damage replace the electrode Instrument defect Contact us for warranty fulfillment Reading keeps jumpi...

Страница 12: ...ons beyond our reasonable control To get the fastest warranty fulfillment go to support aperainst com and click New Support Ticket on the upper right corner Type your email in the requester field Warr...

Отзывы: