Technical Specifications  

ORP (mV) 

Range 

-1000 mV to 1000 mV 

Accuracy 

±0.2% F.S 

Temperature 

Range 

0 to 50˚C (32-122˚F) 

Accuracy 

±0.5˚C 

 

Icons and  Functions  

 

Probe Replacement  

To replace a probe: 1)take off the probe cap; 2) screw off the probe ring 3) unplug the probe; 4) plug in the new 

replacement probe (pay attention to the probe’s position); 5) screw on the probe ring tightly. The replacement probe that 

is compatible with ORP60-Z is 

ORP60-DE (ORP Probe).

 

 

Warrant y  

We warrant this instrument to be free from defects in material and workmanship and agree to repair or replace free of 

charge, at option of APERA INSTRUMENTS, LLC, any malfunctioned or damaged product attributable to responsibility of 

APERA INSTRUMENTS, LLC for a period of TWO YEARS (SIX MONTHS for the probe) from the delivery. This limited 

warranty does NOT cover any damages due to: accidental damage, unauthorized repair, normal wear and tear, or external 

causes such as accidents, abuse, or other actions or events beyond our reasonable control. 

 

To claim for warranty or seek additional help, please go to 

support.aperainst.com

 and click 

New support ticket

. After 

the ticket is submitted, one of our customer care specialists will be assisting you within 1 business day. 

 

APERA INSTRUMENTS, LLC 

Address: 6656 Busch Blvd, Columbus Ohio 43229

Tel: 1-614-285-3080 

Email

[email protected]

 Website: www.aperainst.com 

Self-Diagnosis Symbol 

Er2, Er3 

Stable reading indicator 

 

Waterproof Rating 

IP67, floats on water 

Reading Lock 

HOLD 

Power 

DC3V, AAA batteries*4 

Bluetooth Signal 

 

Battery Life 

200 Hours 

Low power reminder 

           

Backlight 

White: Measurement

Green: 

Calibration

Red: Alarm 

Auto. Power Off 

Automatically power off if no operation for 10 minutes

 

Dimension/Weight 

Instrument

40×40×178mm/133g; case

255×210×50mm/500g; 

Содержание ORP60-Z

Страница 1: ...1 SKU AI3717 ORP60 Z Smart ORP Redox Tester ORP Temp Instruction Manual APERA INSTRUMENTS LLC www aperainst com V1 03...

Страница 2: ...arranty 8 ATTENTION 1 You may find a few drops of water in the probe cap These water droplets are added to maintain the sensitivity of the pH sensor before the product leaves factory It does NOT mean...

Страница 3: ...dial display 3 Graph Mode digital display graph display 4 Table Mode digital display real time measurement and history display Calibration Press buttons to operate Operate on smartphone following gra...

Страница 4: ...n off Bluetooth receiver When turned on will be flashing when connected to smartphone will stay on 2 In parameter setting short press to change parameter Uni directional 1 Long press to enter calibrat...

Страница 5: ...ed e g distilled deionized water for a long time which will make the electrode response slow If this happens soak the ORP electrode in 3M KCl solution for 3 5 hours and then re calibrate it before usi...

Страница 6: ...ound 20 C then adjust the calibration value to 227 mV Notes 1 Usually there s no need to calibrate for ORP testing When the tester has been used for a long time users can test it in 222 mV ORP standar...

Страница 7: ...For example if 10 seconds is set when the measured value is stable for more than 10 seconds the measured value will be automatically locked and the HOLD icon will be displayed Short press to release...

Страница 8: ...ery This limited warranty does NOT cover any damages due to accidental damage unauthorized repair normal wear and tear or external causes such as accidents abuse or other actions or events beyond our...