4) TDS is proportional to the value of conductivity. The TDS coefficient ranges from 0.40 to 

1.00. Users can adjust TDS coefficient by experimental data or experience in P4. The 

following chart lists some commonly used TDS coefficients for reference (Factory default 

value is 0.71). 

 

Conductivity and TDS coefficient 

Conductivity of the solution

 

TDS Coefficient

 

0~100 μS/cm

 

0.60

 

100~1000 μS/cm

 

0.71

 

1~10 mS/cm

 

0.81

 

10~100 mS/cm

 

0.94

 

 

5) Salinity is proportional to the value of conductivity. Its coefficient is 0.5. Users only need 

to calibrate conductity, and then switch to TDS and/or salinity to take measurements. 

 

6) The tester has self-diagnosis functions: 

 

Symbol 

Self-Diagnosis information 

How to fix 

 

 

 

 

Wrong calibration solution, 

which exceeds the 

recognizable range of the 

meter. 

1. Check if calibration solution is 

correct 

2. Check if probe is damaged. 

 

 

 

            is pushed before 

measurement becomes stable 

(        comes up) 

Wait for the smile icon to stay,    and 

then short press           

 

 

 

 

 

 

Содержание COND5

Страница 1: ...COND5 Pocket Conductivity Tester EC TDS Salinity Temp APERA INSTRUMENTS Europe GmbH www aperainst de ...

Страница 2: ...reliable test result and avoid unnecessary damages to the meter or probe For video tutorials please go to www aperainst de Contents 1 Battery Installation 3 2 Keypad Functions 3 3 Complete Kit 4 4 Conductivity Calibration 4 5 Conductivity Measurement 5 6 Notes 5 7 Parameter Setting 7 8 Technical Specifications 8 9 Other Functions and Parameters 8 10 Probe Replacement 9 11 Warranty 9 ...

Страница 3: ...tester and long press to turn off the tester 2 When turned off long press to enter parameter setting 3 In measurement mode short press to turn on backlight 1 In measurement mode short press to switch parameter 2 In mode setting short press to change parameter Unidirectional 1 Long press to enter calibration mode 2 In calibration mode short press to confirm calibration 3 When measured value is lock...

Страница 4: ...on mode Short press to exit 4 Dip the probe into 1413 µS calibration solution Stir gently leave it to stand LCD displays Diagram 3 short press to complete 1 point calibration The tester returns to measuring mode and calibration icon appears at the button left of LCD 5 Rinse probe in distilled water and dry it Follow the steps in 4 3 4 4 to complete 2nd point calibration in 12 88 mS calibration sol...

Страница 5: ... calibration will meet testing requirements 2 The tester has already been calibrated after manufacture Usually users can use the tester right away or test it in the calibration solutions to test its accuracy When error is large calibrate it before using 3 The factory default temperature compensation coefficient is 2 0 C Users can adjust it in P3 parameter setting according to the following table o...

Страница 6: ... 100 1000 μS cm 0 71 1 10 mS cm 0 81 10 100 mS cm 0 94 5 Salinity is proportional to the value of conductivity Its coefficient is 0 5 Users only need to calibrate conductity and then switch to TDS and or salinity to take measurements 6 The tester has self diagnosis functions Symbol Self Diagnosis information How to fix Wrong calibration solution which exceeds the recognizable range of the meter 1 ...

Страница 7: ...inue measuring b Backlight P2 Off turn off the backlight On turn backlight always on 1 backlight lasts 1 minute c Restore to factory default P7 Select Yes to restore the meter s calibration and other parameters back to factory default value When meter s calibration or measuring is performing abnormally users can perform this function After restoring to factory default conduct calibration and measu...

Страница 8: ...e 0 10 00 ppt Resolution 0 01 ppt Temperature Range 0 50 C Resolution 0 1 C Accuracy 0 5 C Screen STN 180 viewing angle Blue backlit measuring mode Green backlit calibration mode Readings Lock HOLD Low battery reminder flashes to remind to replace batteries Automatic power off Auto power off if no operation in 8 minutes Waterproof level IP67 floats on water Power supply AAA batteries 4 Battery Lif...

Страница 9: ...ponsibility of APERA INSTRUMENTS Europe GmbH for a period of two years from the delivery a six month limited warranty applies to electrodes This warranty does not apply to defects resulting from actions such as misuse violation of the instructions in this manual or operations in the manner not specified in this manual improper maintenance and unauthorized repairs Warranty period is the time limit ...

Отзывы: