- 5 -

7. PARAMETERS ADJUSTEMENT

Configuration activities may be performed in the following ways:

1.

 Using USB interface and computer program ARSOFT-WZ1 (on-line):

- connect the recorder to the USB port of computer and run the application
- after connection, the program shows current measurement data, battery voltage and internal time 
  and date of the recorder. "

STATUS

" LED indicator indicates that the transmission is running

- parameters that may be adjusted in the configuration window are presented in table 1
- every change must be confirmed by clicking "

Accept

"

- the program allows to synchronize date and time with the computer
- current configuration values may be stored in a file or restored from a file
- the recorder will update the configuration file (

cfg

) after disconnecting from computer USB (unless 

  battery voltage is < 2.7 V)

     NOTE

:

     - before disconnecting the device from computer, click "

Disconnect the device

"

     - when there is no response: 

- make sure the drivers have been installed properly

   

- disconnect the recorder from the USB interface and reconnect it after a few seconds
- restart the computer
- remove for a few seconds battery from the recorder (section 14)

2.

 

.

 Using configuration file created by ARSOFT-WZ4 (off-line):

    - run the application and in the "

Device

" box enter the name of the device (

AR232

 or 

AR236

)

    - adjust parameters showed according to the table 1 (except for 

RTC

 and 

ID

)

    - current configuration may also be created via changing values from an existing file
    - save the created configuration in a file (

AR236.cfg

 or

 AR232.cfg

) and store it on an SD/MMC card 

      (using the card reader)
    - insert the card in the "

SD/MMC CARD"

 slot and press "

COPY

"; while copying the "

STATUS

      indicator will be illuminated
    - when the "

STATUS

" LED turns off, remove the card from the slot

    NOTE

:

    - for the off-line configuration to be successfully completed, parameters for data protection from 
      unauthorized copying and changing should be set as follows:
      

bloc

=

disabled

Prot

=

disabled

 or when 

Prot

=

enabled

 the 

Pass

 parameter value in the SD/MMC 

      card file should match the value of the 

Pass

 parameter in the recorder

    - pressing "

COPY

" button and successful authorization of the card triggers also the action to move 

      archive „

csv

” files onto this card

!

!

  0

RTC

internal time   

2008.06.01

 ÷ 

2099.12.31

hh:mm:ss

08.06.01

of the recorder 

(

1

)

  

1

Start

recording start point 

(

2

)

2008.06.01

 ÷ 

2099.12.31

hh:mm:ss

08.06.01

  2

Stop

recording end point  

(

2

)

2008.06.01

 ÷ 

2099.12.31

hh:mm:ss

08.06.01

  3

MPer

measure.and record.period 

(

3

)

  

1

0

s

 ÷ 

24h

, interval 10s

10 min

  4

SDest

measurement output storage 

internal

 or 

auto

 (SD/MMC when present)

auto

  5

dot

measurement resolution 

0

 = 1°C and 1%RH,  

1

 = 0,1°C and 0,1%RH 

1

=0,1

  6

coHum

zero offset 

 

-5

.

0

 ÷ 

5.0

 %RH (AR236 only)

0 %RH

for humidity 

(

5

)

  7

coTem

zero offset 

 

 

   

-2

.

0

 ÷ 

2.0

 °C

0 °C

for temperature  

(

5

)

  8

bloc

"

COPY

" button lock 

enabled 

or 

disabled

disabled

  9

Pass

card authorization password

0

 ÷ 

9999

 

 

1111

 10

Prot

card authorization request 

(

6

)

enabled 

or 

disabled

disabled

 11

ID

identification number 

(

7

)

0

 ÷ 

9999

 

0

Table 1. Configuration parameters

  #   Name    Description                           Value and value range                                     factory   custom 

Settings

Содержание AR232

Страница 1: ...9 Warsaw ul Klobucka 8 pav 119 Poland Tel 48 22 853 48 56 853 49 30 607 98 95 Fax 48 22 607 99 50 E mail info apar pl Web site www apar pl TEMPERATURE AND HUMIDITY RECORDER AR236 TEMPERATURE RECORDER AR232 Rev 1 1 0 2010 02 16 ...

Страница 2: ...L PARTS IDENTIFICATION 4 5 CONNECTING TO THE COMPUTER AND DRIVER INSTALLATION 4 6 SOFTWARE INSTALLATION 4 7 PARAMETERS ADJUSTMENT 5 8 OPERATION AND FEATURES OF SD MMC CARD 6 9 VIEWING RECORDED MEASUREMENTS AND EVENTS 7 10 ALERT CONFIGURATION 7 11 LCD DISPLAY OPERATION MODES 8 12 MESSAGES AND ERRORS LED INDICATORS 8 13 IMPORTANT OPERATIONAL NOTES 8 14 REPLACING THE BATTERY 9 The producer reserves t...

Страница 3: ...n SD MMC card n checksum to detect unauthorized changes in the archive n uniquely identifying multiple recorders of the same type by assigning unique identifier ID to each of them n low and high value alerts in band and out of band alert LED indicators n programmable measurement period start and end point of recording and other configuration parameters measured signal zero adjustment COPY button l...

Страница 4: ...lithium battery 3 6 V typeAA R6 2450 mAh SAFTLS14500 Working time on new battery 1 up to 5 years in ambient temperature 20 30 C Nominal operation environment 20 70 C 100 RH without condensation Protection level IP20 Casing on wall material ABS UL94 V0 white Casing dimensions 80 x 80 x 25 mm Working position any Weight 97g with battery 1 working time depends on measurement period whether SD MMC car...

Страница 5: ... and SD MMC storage available after inserting a card into the SD MMC CARD slot In internal memory is available a configuration text file AR236 cfg or AR232 cfg section 7 NOTE Do not disconnect the device from the computer while installing drivers or when detecting the discs after connecting the recorder to the USB interface READ WRITE indicator illuminated 6 SOFTWARE INSTALLATION The CD shipped wi...

Страница 6: ...values from an existing file save the created configuration in a file AR236 cfg or AR232 cfg and store it on an SD MMC card using the card reader insert the card in the SD MMC CARD slot and press COPY while copying the STATUS indicator will be illuminated when the STATUS LED turns off remove the card from the slot NOTE for the off line configuration to be successfully completed parameters for data...

Страница 7: ...n which requires removing batteries for a while Notes 1 parameter available only in ARSOFT WZ1 2 when connected to the USB interface of computer the device does not record data in a file 3 the measurement and recording period start with disconnection from USB 4 see section 11 for details about operating modes of the display 5 parameters enabling offset for measurement eg in case discrepancies betw...

Страница 8: ... 6 LVBAT x xx where x xx is battery voltage V Use ARSOFT WZ3 application to produce and print graphical or textual representation of the output Moreover the application also enables detection of unauthorized changes of the archive Alternatively the csv files may be edited with spreadsheet software such as Microsoft Excel or OpenOffice Calc as well as word processors e g Windows WordPad or Notepad ...

Страница 9: ...rement output recording and the L b A t message on the display additionally a record is created in the archive file section 9 13 IMPORTANT OPERATIONAL NOTES Read this user manual carefully to ensure easy usage of the product For maximum durability of batteries make sure that measurement and recording period parameters are set to the highest possible value above 10 min is recommended parameter 3 MP...

Страница 10: ...ace battery with a new one of correct type section 3 make sure the battery is placed correctly in the slot put the circuit board back in place and close the back cover NOTE During battery replacement the real time clock RTC is cleared and needs to be reset with ARSOFT WZ1 application The other parameters remain unchanged Fig 2 Removing circuit board catches Fig 1 Removing back cover ...

Отзывы: