- 5 -
7. PARAMETERS ADJUSTEMENT
Configuration activities may be performed in the following ways:
1.
Using USB interface and computer program ARSOFT-WZ1 (on-line):
- connect the recorder to the USB port of computer and run the application
- after connection, the program shows current measurement data, battery voltage and internal time
and date of the recorder. "
STATUS
" LED indicator indicates that the transmission is running
- parameters that may be adjusted in the configuration window are presented in table 1
- every change must be confirmed by clicking "
Accept
"
- the program allows to synchronize date and time with the computer
- current configuration values may be stored in a file or restored from a file
- the recorder will update the configuration file (
cfg
) after disconnecting from computer USB (unless
battery voltage is < 2.7 V)
NOTE
:
- before disconnecting the device from computer, click "
Disconnect the device
"
- when there is no response:
- make sure the drivers have been installed properly
- disconnect the recorder from the USB interface and reconnect it after a few seconds
- restart the computer
- remove for a few seconds battery from the recorder (section 14)
2.
.
Using configuration file created by ARSOFT-WZ4 (off-line):
- run the application and in the "
Device
" box enter the name of the device (
AR232
or
AR236
)
- adjust parameters showed according to the table 1 (except for
RTC
and
ID
)
- current configuration may also be created via changing values from an existing file
- save the created configuration in a file (
AR236.cfg
or
AR232.cfg
) and store it on an SD/MMC card
(using the card reader)
- insert the card in the "
SD/MMC CARD"
slot and press "
COPY
"; while copying the "
STATUS
"
indicator will be illuminated
- when the "
STATUS
" LED turns off, remove the card from the slot
NOTE
:
- for the off-line configuration to be successfully completed, parameters for data protection from
unauthorized copying and changing should be set as follows:
bloc
=
disabled
,
Prot
=
disabled
or when
Prot
=
enabled
the
Pass
parameter value in the SD/MMC
card file should match the value of the
Pass
parameter in the recorder
- pressing "
COPY
" button and successful authorization of the card triggers also the action to move
archive „
csv
” files onto this card
!
!
0
RTC
internal time
2008.06.01
÷
2099.12.31
,
hh:mm:ss
08.06.01
of the recorder
(
1
)
1
Start
recording start point
(
2
)
2008.06.01
÷
2099.12.31
,
hh:mm:ss
08.06.01
2
Stop
recording end point
(
2
)
2008.06.01
÷
2099.12.31
,
hh:mm:ss
08.06.01
3
MPer
measure.and record.period
(
3
)
1
0
s
÷
24h
, interval 10s
10 min
4
SDest
measurement output storage
internal
or
auto
(SD/MMC when present)
auto
5
dot
measurement resolution
0
= 1°C and 1%RH,
1
= 0,1°C and 0,1%RH
1
=0,1
6
coHum
zero offset
-5
.
0
÷
5.0
%RH (AR236 only)
0 %RH
for humidity
(
5
)
7
coTem
zero offset
-2
.
0
÷
2.0
°C
0 °C
for temperature
(
5
)
8
bloc
"
COPY
" button lock
enabled
or
disabled
disabled
9
Pass
card authorization password
0
÷
9999
1111
10
Prot
card authorization request
(
6
)
enabled
or
disabled
disabled
11
ID
identification number
(
7
)
0
÷
9999
0
Table 1. Configuration parameters
# Name Description Value and value range factory custom
Settings