   Automatikprodukter

Ozone Detector c/w
Indicator for Temperature/Humidity

O3TH

   Dec.11

Guidelines on How to Measure Ozone

The following information is presented to help users operate their
Ozone Detector  in the most effective and efficient manner.

General

• Ozone is heavier than air and tends to sink.

  Thus detection of leaks from ozone generating equipment should be performed at
   the most appropriate position.

•  Ozone will react and decompose on surfaces such as walls,
    furniture etc.

•   Smell is not a reliable test for the presence or concentration of
    ozone as the odour threshold varies widely between people and
    is affected by local ambient conditions.

Permanent Controller Placement

• The Ozone Detector has been designed to measure
   the ambient concentration of ozone.

   The controller must not be placed directly in an ozone stream.

•  For indoor local area monitoring attach the controller to an inert
   surface with the inlet unobstructed.

•   For leak detection mount the unit near the ozone equipment.

•   Ensure that the controller is protected from excessive water
    splashing, dust, vibration, excessive heat or cold, high
    concentrations of ozone and excessive swings in humidity.

False Readings

• The Ozone Detector has been designed to respond
   selectively to ozone, however other oxidizing gases such as
   chlorine and nitrogen dioxide can generate false readings if they
   are at high concentrations.

    High concentrations of hydrocarbon gases such as vapours of alcohol,
    oils and solvents can reduce and mask the concentration of ozone.

Содержание O3TH

Страница 1: ...Features Real time detection of ambient ozone gas level Full time indicate O3 level Wall mount with LCD display Temperature humidity detection The on off output for control a fan O3 generator or an a...

Страница 2: ...t off power first Simualtaneously depress the 2 clips on either of the sides of the detector gently with your nails or other unsharp tools then move off the face part 2 Mount the detector on the place...

Страница 3: ...range is 10 1000ppb accuracy 1ppb and setting range is 40 1000ppb its maximum measurement limited is 2000ppb 3 Turn on Turn on the monitor by press power for 2 seconds Temperature measurement displays...

Страница 4: ...p on press switch key for about 20 seconds until X10 symbol disappearing It indicates the zero has no any calibration it s still 0 Note Auto zero calibration can be done only when ozone measurement is...

Страница 5: ...tion 1 9600bps 2 14400bps 3 3 19200bps 4 28800bps 5 38400bps RESET Put the DIP1 to ON and press switch key for about 25 seconds until the detector is off then all the setting return to the default Ann...

Страница 6: ...is finished 6 Now go to demarcation point 2 put the controller in sample gas 7 Write down the un twinkling point of X2 and Y2 8 Press switch to the second demarcation point N 2 9 Press tangent down o...

Страница 7: ...key for about 25 seconds until the monitor is off then all the settings return to the default If warm up time is less than 3 months before leaving factory then do emendation as the second row of the...

Страница 8: ...ent Controller Placement The Ozone Detector has been designed to measure the ambient concentration of ozone The controller must not be placed directly in an ozone stream For indoor local area monitori...