Automatikprodukter
Ozone Detector c/w
Indicator for Temperature/Humidity
O3TH
Dec.11
Steps:
1) Energize the controller
2) Put the four DIPs to the upper position to be ON and wait until the digits displayed on the LCD to be stable
3) Demarcate demarcation point 1 (0), write down the un-twinkling point of X1
4) Press switch to the first demarcation point (N=1)
5) Press (tangent down) or (tangent up) to adjust the twinkling figure until it equals to X1.
Then the demarcation of X1 is finished
6) Now go to demarcation point 2, put the controller in sample gas
7) Write down the un-twinkling point of X2 and Y2
8) Press switch to the second demarcation point (N=2)
9) Press (tangent down) or to (tangent up) adjust the twinkling figure until it equals to Y2
10) Press switch, then the following figure starts twinkling
11) Press (tangent down) or (tangent up) to adjust the twinkling figure until it equals to X2.
Then the demarcation of X2 is finished
12) Repeat steps 6-11 to finish the emendation of demarcation point 3 to 10
13) After the emendation to all demarcation points, put the four DIPs to OFF.
6
1000PPB 7
5
4
3
2
1 1023 X (A/D transition)
Y ( ozone measurement)