Operation
- 18 -
For Veterinary use only
Inject detergent into each channel
1-
Block the suction cylinder, air/water cylinder(inapplicable to VET-3512), instrument channel with the
channel plug; block the air/water supply connector with the sealing plug; connect the injection tube to
the endoscope connector; ensure that the distal end of the endoscope is thoroughly immersed in the
detergent.
2-
Use a 50ml syringe to respectively inject detergent into the air/water channel (inapplicable to VET-3512)
and suction cylinder of the endoscope through the injection tube at least 15 times (750ml) (as shown in
figure 4.7.3.2).
3-
Disconnect all channel plugs and the injection tube from the endoscope and immerse them into the
detergent.
VET -6010, VET-9833HD,VET -9830(HD), VET -1335(HD), VET -8028HD,VET -8015(HD),VET-9215(HD)
VET-3512
(Figure 4.7.3.2
)
4