ANYLOAD A2A-D2 Amplifier Product Manual (V1810)

 

 

 

 
 
 

5.2 Zero Calibration 

 

1. Remove weights from load cell or scale like hooks, plate, tare, etc 

 

2. If current signal output is desired, set the output current  to 4.00mA  

             by djusting the potentiometer ZERO 
 

    

 

     If voltage signal output is desired, set the output voltage to 10.00V  

             by adjusting the potentiometer ZERO 
 

Note1: When adjusting the potentiometer to set zero, slightly rotate  
clockwise or counter clockwise the potentiometer’s screw. For higher  
accuracy output, it is recommended to set zero to 4.00mA or 10.00V 
 
Note2: The tare weight shall not exceed 20% full scale capacity of scale 
or load cell  
 
Note3: When it gives output a complimentary signal (negative voltage), 
remove the load first then do zero calibration.  

 

5.3 Span Calibration 

 

1. Apply load up to the max capacity of the load cell or scale ( max 

 

load can be at least 80% of the max capacity of load cell or scale) 

 

2. If current signal output is desired, set the output current  to  

 

    20.00mA by adjusting the potentiometer SPAN 

 

     If voltage signal output is desired, set the output voltage to  
     10.00V (-10.00v) or 5.00V (-5.00V) by adjusting the potentiometer  
     SPAN 

 

 

 

If the gain range is not suitable after the above 2 steps, change the 

 

Gain range ( 4-bit DIP switch ) and recalibrate 

 

Note1: For higher accuracy of SPAN output, it is recommended to repeat  
steps in Zero Calibration and then Span calibration ( steps 5.2 & 5.3 )  
multiple times until you reach your desired output 
 
Note2: Potentiometers IN1 & IN2 shall be adjusted if the amplifier is 
applied of having multiple input load cells or scales. When multiple load 
cells are connected, the maximum difference in sensitivity between each 
load cell shall not exceed 3%.

 

 
 

Содержание A2A-D2

Страница 1: ...Anyload Transducer Co Ltd Website www anyload com Email info anyload com Fax 1 866 612 9088 North America Toll Free 1 855 ANYLOAD 269 5623 A2A D2 V1611 Load Cell Amplifier Product Manual V1810...

Страница 2: ...D2 Amplifier Product Manual V1810 TABLE OF CONTENTS 1 Introduction 2 2 Installation 2 3 Specifications 2 4 Connection Switch Diagrams 3 5 Operations 4 6 Maintenance 6 7 Warnings and Precautions 6 8 Il...

Страница 3: ...ped with individual adjustable resistors and has a wide input signal of 0 36mV and output signal of 0 10V or 2 to 20mA The manual here provides the installation operation and calibration procedures of...

Страница 4: ...y the voltage terminals output can give complimentary voltage output 10V 4 3 Excitation voltage Through the slide switch EXC the A2A D2 can provide either 12VDC or 5VDC excitation voltage to the load...

Страница 5: ...n switches either 2 3 or 4 Table 4 4 2 Gain Switch Settings 4 bit DIP Switch No of DIP Switch Gain Range GAIN 1 217 753 2 153 303 3 103 153 4 86 118 Note If the Gain setting is not achieved during cal...

Страница 6: ...libration 5 3 Span Calibration 1 Apply load up to the max capacity of the load cell or scale max load can be at least 80 of the max capacity of load cell or scale 2 If current signal output is desired...

Страница 7: ...enced personnel 3 The amplifier can only support up to 8 load cells with 350 Ohms each or combination of multiple load cells with equivalent power consumption for 8x350 Ohms load cells 7 Warnings Prec...

Страница 8: ...GND Ground EXC Switch of excitation ZERO Zero potentiometer SPAN Span potentiometer Io Positive current output Common positive Vo Positive voltage output GAIN Gain selection IN 1 Load cell 1 connectio...