40 DECADE II user manual, ed. 11
Fig. 18. Example of a signal with regular evenly spaced alterations: a sine.
Fig. 19. Sine of Fig. 18. A full period is 0.41 min (25 s) which corresponds to
a frequency of 1/25 = 0.04 Hz.
Frequency of signal and noise
Also a chromatographic peak can be expressed in terms of frequencies. The
way to determine this frequency is the same. The duration of the full peak is
measured and expressed in Hz.
Fig. 20. Frequency tells how often something happens: 1 peak in about 0.25
min (15 s), f = 1/15 = 0.07 Hz.