Reflection Magnitude Uncertainty
37x47D/3652/Reflection Only
U
n
c
e
rt
a
in
ty
(
d
B
)
10
1
0.1
-40
-35
-30
-25
-20
-15
-10
-5
0
0.04 GHz
2 GHz
20 GHz
Device S11 (dB)
37X47D (40 MHz – 20 GHz)
System Dynamic Range
is defined as the difference between the available power from Port 1 incident on Port 2 in a through line connection and
the noise floor.
VNA settings:
10 Hz I.F. Bandwidth, 101 pts, 12-term SOLT calibration, 512 averages
Corrected Specifications
Test Port Characteristics
3
Connector
Frequency
1
(GHz)
Directivity
(dB)
Source Match
(dB)
Load Match
(dB)
Reflection Tracking
(dB)
Transmission Tracking
(dB)
Isolation
(dB)
K (2.92 mm)
0.04
>42
>40
>42
±0.005
±0.030
>105
20
>42
>38
>42
±0.006
±0.070
>110
The graphs give measurement uncertainty after 12-Term vector error correction. The errors are worst case contributions of residual directivity,
load and source match, frequency response, isolation, network analyzer dynamic accuracy, and connector repeatability.
Reflection Phase Uncertainty
37x47D/3652/Reflection Only
Device S11 (dB)
U
n
c
e
rt
a
in
ty
(
D
e
g
re
e
s
)
100
10
1
-40
-35
-30
-25
-20
-15
-10
-5
0
0.04 GHz
2 GHz
20 GHz
U
n
c
e
rt
a
in
ty
(
d
B
)
Device S21 (dB)
Transmission Magnitude Uncertainty
37x47D/3652/Transmission Only
10
1
0.1
0.01
-70
-60
-50
-40
-30
-20
-10
0
10
-80
0.04 GHz
2 GHz
20 GHz
0.04 GHz
2 GHz
20 GHz
Transmission Phase Uncertainty
37x47D/3652/Transmission Only
U
n
c
e
rt
a
in
ty
(
D
e
g
re
e
s
)
100
10
1
0.1
Device S21 (dB)
-80
-70
-60
-50
-40
-30
-20
-10
0
10
Reflection Measurements
Transmission Measurements
Model
Frequency
1
(GHz)
Port 1 Power, Typical (dBm)
Noise Floor at Port 2 (dBm)
2
System Dynamic Range (dB)
37347D
0.04
10
–82
92
2
11
–104
115
20
8
–100
108
37247D
0.04
10
–82
92
2
11
–104
115
20
7
–101
108
Measurement Uncertainty
10
11
7
10
11
6
With Option 15
–82
–104
–99
–82
–104
–100
With Option 15
With Option 15
92
115
106
92
115
106
2
>42
>40
>42
±0.005
±0.030
>105
NOTES: 1. Specifications between the stated frequencies are to be linearly interpolated as typical values.
2. RMS values – compensated for loss in available source power from test port cable connected between Port 1 and Port 2.
3. The specifications for Test Port Characteristics apply when the VNA Universal Test Port Adapters are connected, with or without
phase equal insertables, to the test set ports and calibrated with the appropriate calibration kit at 23°C ±3°C using the SOLT
calibration method with a sliding load to achieve 12-term error correction. (90 min. warm-up time is recommended.)