Anrtisu Lightning 37000D Скачать руководство пользователя страница 1

Lightning

37000D
Vector Network Analyzers

40 MHz to 65 GHz

Technical Datasheet and Configuration Guide

37000D

Series

Network Analysis Solutions for Design and Manufacturing

 

 

 

 

 

 

 

 

 

 

Содержание Lightning 37000D

Страница 1: ...Lightning 37000D Vector Network Analyzers 40 MHz to 65 GHz Technical Datasheet and Configuration Guide 37000D Series Network Analysis Solutions for Design and Manufacturing...

Страница 2: ...000 1 Data Averaging Point by point default up to 4096 Sweep by Sweep IF Bandwidth 10 kHz 10 Hz Reference Plane Can be entered in time or in distance when the dielectric constant is entered Automatic...

Страница 3: ...ing an external power meter over the Dedicated GPIB port Multiple Source Control Capability Multiple Source Control capability allows a user to independently control the frequencies of two sources and...

Страница 4: ...64 bit floating point data can be transferred with LSB or MSB first Measurement Throughput Measurement times are based on a single 40 MHz to 20 GHz sweep with 10 kHz IF bandwidth no averages after a...

Страница 5: ...Interfaces Front Panel Connectors and Controls Keyboard Input Any PS 2 keyboard can be connected to the front panel for navigating through front panel menus annotation of data files and display label...

Страница 6: ...provide a range of trade offs of main lobe width versus side lobe level ringing The general type of function used is the Blackman Harris window with the number of terms being varied from one to four T...

Страница 7: ...e S21 dB Transmission Magnitude Uncertainty 37x47D 3652 Transmission Only 10 1 0 1 0 01 70 60 50 40 30 20 10 0 10 80 0 04 GHz 2 GHz 20 GHz 0 04 GHz 2 GHz 20 GHz Transmission Phase Uncertainty 37x47D 3...

Страница 8: ...Power Typical dBm Noise Floor at Port 2 dBm 2 System Dynamic Range dB 37369D 0 04 10 85 95 2 8 107 115 20 3 103 106 40 2 95 97 37269D 0 04 10 85 95 2 8 107 115 20 2 104 106 40 2 97 99 The graphs give...

Страница 9: ...z Port 1 Power Typical dBm Noise Floor at Port 2 dBm 2 System Dynamic Range dB 37377D 0 04 10 88 98 2 5 110 115 20 2 106 108 40 1 98 99 50 1 94 93 37277D 0 04 10 88 98 2 5 110 115 20 1 107 108 40 1 10...

Страница 10: ...ncy1 GHz Port 1 Power Typical dBm Noise Floor at Port 2 dBm 2 System Dynamic Range dB 37397D 0 04 10 88 98 2 5 110 115 20 2 106 108 40 1 98 99 50 1 94 93 65 2 82 80 37297D 0 04 10 88 98 2 5 110 115 20...

Страница 11: ...els the port 1 power may be attenuated in 10 dB steps using the internal 70 dB 60 dB for 37377D and 37397D step attenuator Similarly high input signals into port 2 not exceeding 1 watt can be attenuat...

Страница 12: ...Ta b2 Tb Source Transfer Switch ALC Port 1 ALC Port 2 Sampler Buffer Amplifier Coupler Coupler Port 1 Port 2 a1 Ra a2 Rb b1 Ta b2 Tb Input Output External Source Loop Input Output External Source Loo...

Страница 13: ...Port 2 Step Attenuator Step Attenuator b2 In b2 Out 37300D Block Diagram 37300D Block Diagram with Option 15 13 General Power Requirements 85 240 volts 48 63 Hz 540 VA maximum Dimensions 267 H x 432 W...

Страница 14: ...Calibration Calibration Kits Using Anritsu s precision calibration kits ensures accurate operation of your 37000D series VNA These kits include precision components required to perform full 12 term c...

Страница 15: ...o Coax Adapter 33 to 50 GHz WR22 V m n n 35WR22VF Waveguide to Coax Adapter 33 to 50 GHz WR22 V f n n 35WR28K Waveguide to Coax Adapter 26 5 to 40 GHz WR28 K m n n 35WR28KF Waveguide to Coax Adapter 2...

Страница 16: ...o K f n n 43KB 20 Fixed Attenuator 20 dB DC to 26 5 GHz 50 K m to K f n n 43KC 3 Fixed Attenuator 3 dB DC to 40 GHz 50 K m to K f n n 43KC 6 Fixed Attenuator 6 dB DC to 40 GHz 50 K m to K f n n 43KC 1...