Chapter 5 — Component Replacement
5-9 Low Band Switch
MN469xC Series Multiport Test Set MM
PN: 10410-00730 Rev. D
5-11
A6 Replacement
1.
Remove the top cover from the test set as instructed in
2.
Replace the A6 module as illustrated in
.
Note
Use Anritsu 01-201 Torque Wrench to tighten all RF connectors when installing the module.
1. Disconnect the RF cable from the J1 connector of the
A6 module and the Rear Panel.
2. Remove the two phillips mounting screws from the
Module.
3. Disconnect the two RF cables from J2 and J3
connectors, then disconnect the control cable from
the module side and remove the module.
4. Installation is the reverse of removal.
Figure 5-10.
A6 Low Band Switch Replacement
1
2
3
A6
J1
J2
J3