Chapter 3 — Performance Verification
3-6 Directivity and Test Port Match Verification
MN469xC Series Multiport Test Set MM
PN: 10410-00730 Rev. D
3-9
28.
Select
Response
|
Tr2
|
S22
29.
Select
Tr3
|
More Single Mode
|
S33
30.
Select
Tr4
|
More Single Mode
|
S44
31.
Insert the Calibration Kit Components Coefficients USB Memory Stick into one of the front panel USB
ports.
32.
Select
Calibration
|
Cal Kit/AutoCal Characterization
|
Install Kit/Charac.
33.
Select
Cal Kit
file type and then click the
Browse
button. See
below.
34.
Select the file with “.ccf” file extension and then click the
Open
button. See
35.
Click on the
OK
button and then the
Install
button to install the Calibration Kit Components Coefficients.
36.
Select
Calibration
|
Calibrate
|
Manual Cal
|
1-Port Cal
|
Modify Cal Setup
|
Edit Cal Params
Figure 3-1.
Install Dialog Box
Figure 3-2.
Open Dialog Box