Model OSL Series CG
PN: 10100-00055 Rev. B
2-7
Performance Verification
2-3
Verification Procedures
Precision Load DC Resistance Verification
Equipment Required
• Agilent Model 34401A Multi-meter
Procedure
Use an Ohmmeter to measure the DC resistance between the outer
conductor and center conductor of the Precision Load and verify that it
is within specification (refer to
Table 2-1, “OSL Series Specifications
).