Anritsu MT8870A Скачать руководство пользователя страница 17

17 

 

 

4)

 

Click the measurement execution button (triangle symbol) at the top of the screen to perform 
measurement.

 

Check the displayed graph and confirm that the waveform has a peak.

 

 

This confirms RF signal output and measurement.

 

 

If the waveform does not have a peak, check the following:

 

- RF cables connected to correct port set at VSA/VSG Port in SRW Applet screen

 

- Power and Frequency settings of CW Analysis match VSG settings

 

- Modulation of VSG set to CW

 

- RF signal output started (Click 

Transmit

 at the VSG screen to check this.)

 

 

The method for checking performance is described in Chapter 7 of the PDF document labeled 
mu887000a_opm_e_xxx.pdf.

 

 

 

Содержание MT8870A

Страница 1: ...Quick Start Guide Universal Wireless Test Set MT8870A...

Страница 2: ...Utility Tool 7 2 7 Changing Network Settings 8 2 8 Changing GPIB settings 9 2 9 Upgrading Firmware 10 2 10 Installing Waveform File 12 2 11 Restoring PC Network Configuration 12 3 Checking Operation 1...

Страница 3: ...cables supported DVD MT8870A Universal Wireless Test Set Supplied accessory Useful if you have RF Cable RF cable with N type connector both ends Used in Chapter 3 Checking Operations 1 3 Terms The ter...

Страница 4: ...to read DVD R Memory 1 GB or more HDD Free Space 200 MB or more Display resolution of 1024 768 pixels or better PDF Viewer Interface 1 Ethernet or GPIB 1 An Ethernet connection is recommended for ini...

Страница 5: ...test Utility Tool from the Anritsu Web site as described in Chapter 4 2 4 Changing PC Network Configuration The Utility Tool can control the MT8870A via both GPIB and Ethernet but functions requiring...

Страница 6: ...2 168 1 x where x is 1 to 4 change the IP address to 192 168 1 100 to match the network configuration Note In this description the PC IP address is set to 192 168 1 100 but any IP address in the range...

Страница 7: ...cable The MT8870A has Ethernet connectors at front and back Use either one 2 6 Starting Utility Tool Double click the Utility Tool icon on the PC desktop Click IPv4 to display the Network interface s...

Страница 8: ...as described below Click Configure Network at the Select Interface Function Tree 1 Change the settings to match the network environment 2 Click Apply to set the new network settings at the MT8870A an...

Страница 9: ...to control the MT8870A over GPIB 1 Select the chassis from the Select Interface Function Tree and click Maintenance Settings 2 Set the GPIB address and Delimiter Terminator for each module 3 Click Ap...

Страница 10: ...directory containing the new firmware 3 Select the firmware file 4 Click Upgrade 5 The Progress screen displays the upgrade execution status MU887000A status indicator lamps 1 to 6 blink during the up...

Страница 11: ...cific modules 1 Select the chassis and click Upgrade Firmware 2 Click Browse to select the directory containing the new firmware 3 Select the firmware file 4 Check un check the module s for firmware u...

Страница 12: ...Files 2 Click Browse to select the directory containing the waveform file 3 Select the waveform file 4 Select the destination module 5 Click Transfer Note The waveform file installed at factory shipm...

Страница 13: ...installers in the folder MT8870A Installer CombiView on the accessory DVD See Chapter 2 in the PDF document mx880052a_opm_e_xxx pdf in the folder MT8870A Manual English for the installation instructio...

Страница 14: ...are is described below 1 At the software selection screen double click Cellular Common Spectrum Monitor to display the Spectrum Monitor screen 2 Set the parameters shown below Next configure the Vecto...

Страница 15: ...rm does not have a peak check the following RF cables connected to correct port set at Input Port Output Port at Spectrum Monitor screen Level and Frequency settings of Spectrum Monitor match settings...

Страница 16: ...en double click Short Range Wireless Bluetooth WLAN to start the SRW Applet screen 2 Select Configure VSG from the menu bar to display the Configure VSG screen 3 Set the parameters as shown below If t...

Страница 17: ...ement If the waveform does not have a peak check the following RF cables connected to correct port set at VSA VSG Port in SRW Applet screen Power and Frequency settings of CW Analysis match VSG settin...

Страница 18: ...manuals by registering the user and product at the Anritsu Web site A notification email is sent when software and operation manuals are updated 4 1 If You Do Not Have Anritsu ID First register the u...

Страница 19: ...19 3 Input the required information and click NEXT...

Страница 20: ...20 4 Input the 6 digit confirmation code sent to your email address and click Submit...

Страница 21: ...21...

Страница 22: ...ribed in section 4 1 Go to http www anritsu com and follow the steps listed below to register the purchased MT8870A Log in to My Anritsu and Select Register My Product Input the Model Number and Seria...

Страница 23: ...23 5 Enter the serial number of the MT8870A you purchased and click ADD Once the serial number is confirmed click Submit to complete registration 2 1 3...

Страница 24: ...te Navigate to the software download page shown below 1 Visit http www anritsu com 2 Log in to My Anritsu 3 Select MY PRODUCTS 4 Select MT8870A as the filter condition and click Update 5 Click MT8870A...

Страница 25: ...ration for details Product registration may take a few days to confirm the details Trademarks and Registered Trademarks Microsoft and Windows are registered trademarks of Microsoft Corporation in the...

Страница 26: ...hone 43 0 1 717 28 710 United Arab Emirates Anritsu EMEA Ltd Anritsu A S 2106 Singapore Anritsu Pte Ltd 11 Chang Charn Road 04 01 Shriro House Singapore 159640 Phone 65 6282 2400 Fax 65 6282 2533 Viet...

Отзывы: