15
3)
Click the measurement execution button (triangle symbol) at the top of the screen to perform
measurement.
Check the displayed graph and confirm that the waveform has a peak.
This confirms RF signal output and measurement.
If the waveform does not have a peak, check the following:
- RF cables connected to correct port set at Input Port/Output Port at Spectrum Monitor screen
- Level and Frequency settings of Spectrum Monitor match settings of Vector Signal Generator
- Modulation of Vector Signal Generator set to Off
- Output of Vector Signal Generator set to On
The method for checking performance is described in Chapter 7 of the PDF document labeled
mu887000a_opm_e_xxx.pdf.