MS2830A-077 Analysis Bandwidth Extension to 62.5 MHz (Requires MS2830A-005 and MS2830A-006)
MS2830A-078 Analysis Bandwidth Extension to 125 MHz (Requires MS2830A-005, MS2830A-006 and MS2830A-077)
An image response is received when setting the bandwidth to more than 31.25 MHz. This can be used when not inputting a signal frequency
outside the MS2830A analysis bandwidth (125 MHz max.). The MS2690A/91A/92A Signal Analyzer series is recommended for other
measurement purposes.
General
Analysis bandwidth
Sets capture analysis bandwidth from center frequency
1 kHz to 25 MHz (1-2.5-5 sequence), 31.25 MHz, 50 MHz, 62.5 MHz (with MS2830A-077)
1 kHz to 25 MHz (1-2.5-5 sequence), 31.25 MHz, 50 MHz, 62.5 MHz, 100 MHz, 125 MHz (with MS2830A-078)
Sampling rate
Auto setting by conditions of analysis bandwidth
2 kHz to 100 MHz (1-2-5 sequence) (with MS2830A-077)
2 kHz to 200 MHz (1-2-5 sequence) (with MS2830A-078)
Capture time
with MS2830A-077, >31.25 MHz bandwidth
Setting capture time length
Minimum capture time length: 1 μs
Maximum capture time length: 500 ms
Setting mode: Auto, Manual
with MS2830A-078, >31.25 MHz bandwidth
Setting capture time length
Minimum capture time length: 500 ns to 1 μs (Determined according to analysis bandwidth)
Maximum capture time length: 500 ms
Setting mode: Auto, Manual
ADC resolution
with MS2830A-077/078, >31.25 MHz bandwidth
14 bits
Frequency
Frequency setting
with MS2830A-077/078, >31.25 MHz bandwidth
300 MHz to 3.6 GHz [MS2830A-040]
300 MHz to 6 GHz [MS2830A-041]
300 MHz to 13.5 GHz [MS2830A-043]
Resolution bandwidth
(RBW)
with MS2830A-077, >31.25 MHz bandwidth
Setting range: 1 Hz to 3 MHz (1-3 sequence)
Selectivity (–60 dB/–3 dB): 4.5:1 (nominal)
with MS2830A-078, >31.25 MHz bandwidth
Setting range: 1 Hz to 10 MHz (1-3 sequence)
Selectivity (–60 dB/–3 dB): 4.5:1 (nominal)
44
Product Brochure
l
MS2830A
Содержание MS2830A-040
Страница 51: ...Note Product Brochure l MS2830A 51...