Страница 1: ...Connectors for the MS4642A B or MS4644A B VNA VectorStar MN4697C V Connectors for the MS4645A B or MS4647A B VNA Part Number 10410 00738 Revision C Published June 2017 Copyright 2017 Anritsu Company...
Страница 2: ...Preparation 1 If mounting on a workbench surface first position the MN469xC Multiport Test Set with access to its front and rear panels Place the VNA on top 2 If mounting into rack or console make sur...
Страница 3: ...ure that the airflow hole pattern at the right side of the instrument is clear This is necessary to provide adequate ventilation for the test set Note When front panel loops on a VectorStar MS464xA B...
Страница 4: ...nnector as shown in Figure 2 2 On the VNA rear panel remove and set aside the eight SMA m m loops 3 Install the eight semi rigid cables provided as shown in Figure 2 and Table 1 4 Connect the AC Power...
Страница 5: ...Rear Panel Connections 2 3 62112 81 SMA male male semi rigid Tighten using an 8 mm 5 16 in torque end wrench set to 0 9 N m 8 lbf in MS464xB port labeled b2 loop out MN469xC port labeled TO VNA b2 OUT...
Страница 6: ...ue end wrench set to 0 9 N m 8 lbf in MS464xB port labeled P1 Source loop in MN469xC port labeled TO VNA Port 1 Src INPUT 10 2100 1 Rear Panel GPIB Cable 1 meter 39 3 long IEEE 488 2 GPIB Dedicated GP...
Страница 7: ...test set the VNA application will stay in 2 port mode and only 2 port mode features and functions will be available 4 If the 4 port functions fail to appear exit the VNA application by selecting File...
Страница 8: ...Anritsu Company 490 Jarvis Drive Morgan Hill CA 95037 2809 USA http www anritsu com Anritsu utilizes recycled paper and environmentally conscious inks and toner...