3. Rear Panel DIP Switch GPIB Address Setting

VectorStar MN469xC QSG

PN: 10410-00738  Rev. C

QSG-7

3.

Rear Panel DIP Switch GPIB Address Setting

The MN469xC Series Test Set GPIB address must match the GPIB address set on the VNA and is set on the 
Test Set by rear panel DIP switches. The factory default GPIB address is 

16

 (where

 Switch 1 = ON 

or up and all 

other switches 

OFF 

or down). The GPIB address on the VNA can be verified by navigating to the 

REMOTE 

INTER

 menu and the 

Multiport Test Set

 button.

• Navigation: 

Main Menu | System | Remote Interface | REMOTE INTER | Multiport Test Set

.  

4.

Power Up Sequence

1. Make sure all of the cables are properly installed as illustrated in the previous sections.
2. Turn on the Test Set prior to launching the VNA application on the VNA. The VNA can be powered on, 

but the application must be off.

3. Once the Test Set is on, launch the VNA application. During the launch sequence, the VNA application 

recognizes the Test Set on the GPIB bus and is configured for 4-port mode. If the VNA application is 
powered up and launched before the test set, the VNA application will stay in 2-port mode and only 
2-port mode features and functions will be available.

4. If the 4-port functions fail to appear, exit the VNA application by selecting 

File | Exit 

from the VectorStar 

Menu Bar and then clicking 

Yes

 in the confirmation dialog box.

• Navigation: 

MENU BAR

 | File | FILE Drop-Down Menu | Exit Command

5. After the Windows desktop appears, launch the VNA application by doing one of the following:

a. On the desktop, click the 

VectorStar

 icon. If the VNA was running in 100,000 point mode, the icon 

is annotated with “100K”. 

b. If running in 25,000 point mode, select 

Start

 | All Programs | VectorStar | VectorStar

.

c. If running in 100,000 point mode, select 

Start

 | All Programs | VectorStar_100K | VectorStar.

Figure 3.

MN469xC DIP Switches (factory setting)

Note

The VNA GPIB connection must be made to the 

VNA Dedicated GPIB 

connector and 

NOT

 to the 

VNA IEEE-488.2 GPIB

 connector. 

Note

The VNA application must be started after the Test Set is connected and powered up. If the VNA 
application is started before the Test Set, the VNA will remain in 2-port mode and the 4-port functions 
will not be available. If this happens, exit the VNA application, make sure the Test Set is powered up, 
and launch the VNA application.

1 6   8   4   2   1   T   L   S

Содержание MN469xC Series

Страница 1: ...Connectors for the MS4642A B or MS4644A B VNA VectorStar MN4697C V Connectors for the MS4645A B or MS4647A B VNA Part Number 10410 00738 Revision C Published June 2017 Copyright 2017 Anritsu Company...

Страница 2: ...Preparation 1 If mounting on a workbench surface first position the MN469xC Multiport Test Set with access to its front and rear panels Place the VNA on top 2 If mounting into rack or console make sur...

Страница 3: ...ure that the airflow hole pattern at the right side of the instrument is clear This is necessary to provide adequate ventilation for the test set Note When front panel loops on a VectorStar MS464xA B...

Страница 4: ...nnector as shown in Figure 2 2 On the VNA rear panel remove and set aside the eight SMA m m loops 3 Install the eight semi rigid cables provided as shown in Figure 2 and Table 1 4 Connect the AC Power...

Страница 5: ...Rear Panel Connections 2 3 62112 81 SMA male male semi rigid Tighten using an 8 mm 5 16 in torque end wrench set to 0 9 N m 8 lbf in MS464xB port labeled b2 loop out MN469xC port labeled TO VNA b2 OUT...

Страница 6: ...ue end wrench set to 0 9 N m 8 lbf in MS464xB port labeled P1 Source loop in MN469xC port labeled TO VNA Port 1 Src INPUT 10 2100 1 Rear Panel GPIB Cable 1 meter 39 3 long IEEE 488 2 GPIB Dedicated GP...

Страница 7: ...test set the VNA application will stay in 2 port mode and only 2 port mode features and functions will be available 4 If the 4 port functions fail to appear exit the VNA application by selecting File...

Страница 8: ...Anritsu Company 490 Jarvis Drive Morgan Hill CA 95037 2809 USA http www anritsu com Anritsu utilizes recycled paper and environmentally conscious inks and toner...

Отзывы: