: PATTern: NAME: DEFined1 <Character>

: PATTern: NAME: DEFined2 <Character>

: PATTern: NAME: DEFined3 <Character>

: PATTern: NAME: DEFined4 <Character>

Function ............ The Pattern Generator free pattern memory Defined: 1 to 4 name is set.

Parameter .......... <Character> =

Maximum 8 alphanumeric characters

Unit ................... <Non term>

Restriction ......... None

: PATTern: NAME: DEFined1?

: PATTern: NAME: DEFined2?

: PATTern: NAME: DEFined3?

: PATTern: NAME: DEFined4?

Function ............ The Pattern Generator free pattern memory Defined: 1 to 4 name is inquired.

Response ........... The Pattern Generator free pattern memory Defined: 1 to 4 name

Restriction ......... None

: PATTern: TOP: DEFined1 <Numeric>

: PATTern: TOP: DEFined2 <Numeric>

: PATTern: TOP: DEFined3 <Numeric>

: PATTern: TOP: DEFined4 <Numeric>

Function ............ The first delivery address of the Pattern Generator free pattern memory Defined: 1 to 4 is set.

Parameter .......... <Numeric> =

00000 to 65535

 (a decimal number)

Unit ................... <Non term>

Restriction ......... This cannot be set during the data pattern delivery, but can be set in the idling.

: PATTern: TOP: IDLE <Numeric>

Function ............ The first delivery address of the Pattern Generator idle pattern memory is set.

Parameter .......... <Numeric> =

00000 to 65535

 (a decimal number)

Unit ................... <Non term>

Restriction ......... This cannot be set in the Pattern Generator output ON status.

: PATTern: TOP: DEFiend1?

: PATTern: TOP: DEFiend2?

: PATTern: TOP: DEFiend3?

: PATTern: TOP: DEFiend4?

Function ............ The first delivery address of the free pattern memory Defined: 1 to 4 is inquired.

Response ........... The first delivery address of the free pattern memory Defined: 1 to 4

Restriction ......... None

: PATTern: TOP: IDLE?

Function ............ The first delivery address of the idle pattern memory is inquired.

Response ........... The first delivery address of the idle pattern memory

Restriction ......... None

5.3

Command Details

Содержание MG3641A

Страница 1: ...Fifth Edition For safety and warning information please read this manual before attempting to use the equipment Additional safety and warning information is provided within the MG3641A MG3642A Synthes...

Страница 2: ...ing safety symbols are used inside or on the equipment near operation locations to provide information about safety items and operation precautions Ensure that you clearly understand the meanings of t...

Страница 3: ...product s country of origin for re export from your country Before re exporting the product or manuals please contact us to confirm whether they are export controlled items or not When you dispose of...

Страница 4: ...EU CE marking 1 Product Model Option MG3641A MG3642A 23 Pattern Generator 2 Applied Directive and Standards When the MG3641A MG3642A 23 Pattern Generator is installed in the MG3641A MG3642A the appli...

Страница 5: ...g 1 Product Model Option MG3641A MG3642A 23 Pattern Generator 2 Applied Directive and Standards When the MG3641A MG3642A 23 Pattern Generator is installed in the MG3641A MG3642A the applied directive...

Страница 6: ...vi...

Страница 7: ...Preparation Before Use 3 2 1 Data and Clock Connectors Labeling 3 2 2 Connection 4 Section 3 Operation Overview 7 Section 4 Detailed Description 9 4 1 Delivery Mode 9 4 2 Data Output Method 13 Sectio...

Страница 8: ...II...

Страница 9: ...requency Besides when no data pattern is delivered the idle pattern can be delivered Various FSK modulated waves can be output by combining with the FSK encoder which is optionally built in MG3641A MG...

Страница 10: ...00000 to 65535 Data bit length setting range 2 to 524288 bit However the last delivery address shall be 65535 Write by 1 byte unit via GPIB Write enabled when the Pattern Generator output is turned OF...

Страница 11: ...the Int Mod Cont 1 2 and 3 connectors on the MG3641A MG3642A rear panel Each connector is allocated as follows Int Mod Cont 2 3 1 Data 21 Data 20 Clock MG3641A MG3642A Rear Panel By sticking the attac...

Страница 12: ...plemented the Int Mod Cont connections are as follow DATA PATTERN GENERATOR Output ON OFF Data 21 Output Data 20 Output Clock Output Data 21 Input Data 20 Input Clock Input FSK Encoder ON OFF FSK ENCO...

Страница 13: ...lision by turning OFF the Pattern Generator output when the FSK modulation is performed by using the external data generator etc Also read the FSK Encoder Operation Manual as well if FSK modulation is...

Страница 14: ...6 Section 2 Preparation Before Use...

Страница 15: ...It cannot be changed during the data pattern delivery Send Mode Send mode is selected from the one time send Single or the repeating send Continuous by using the Sel F1 key It cannot be changed durin...

Страница 16: ...set to ON And the free pattern Defined 1 to 4 can display each name which are consist of maximum eight letters The name is written by the GPIB interface Rtn Configuration 1 TEST 1 2 TEST 2 3 TEST 3 4...

Страница 17: ...pattern delivery is finished In the Single mode after the free pattern is delivered once the idle pattern delivery is resumed In the Continuous mode the free pattern delivery is repeated until the sto...

Страница 18: ...n is delivered twice the idle pattern delivery is resumed In the Continuous mode the PN9 pseudo random pattern delivery is repeated until the stop operation is performed When the stop operation is per...

Страница 19: ...idle pattern delivery is resumed In the Continuous mode the PN15 pseudo random pattern delivery is repeated until the stop operation is performed When the stop operation is performed the idle pattern...

Страница 20: ...tern is delivered once the idle pattern delivery is resumed In the Continuous mode the 01 repetition pattern delivery is repeated until the stop operation is performed When the stop operation is perfo...

Страница 21: ...tput starting from the specified first delivery address LSB for the specified data bit length taken out from the Data 21 Output Address D7 MSB D6 D5 D4 D3 D2 D1 D0 LSB 65535 65534 65533 00004 00003 00...

Страница 22: ...65534 65533 00004 00003 00002 00001 00000 2 Rate D0 D2 D4 D1 D3 D5 Int Mod Cont 2 3 1 Data 21 Data 20 Clock Note When the specified data bit length is an odd number the delivery is made of which bit l...

Страница 23: ...ftware specifications comply IEEE488 2 and SCPI Standard Commands for Programmable Instruments GPIB command tree and the command detailed description of Pattern Generator is described in this chapter...

Страница 24: ...DEFined4 DEFined4 IDLE IDLE DEFined1 DEFined1 DEFined2 DEFined2 DEFined3 DEFined3 DEFined4 DEFined4 IDLE IDLE DEFined1 DEFined2 DEFined3 DEFined4 IDLE ERASe ERASe ERASe ERASe ERASe Character Character...

Страница 25: ...t ON OFF is inquired Response The Pattern Generator status OFF ON Restriction None PATTern SELect Character Function The Pattern Generator delivery data pattern is selected Parameter Character DEF1 DE...

Страница 26: ...bps Restriction None PATTern RATE Function The Pattern Generator delivery rate is inquired Response The Pattern Generator delivery rate Restriction None PATTern TYPE Character Function The Pattern Gen...

Страница 27: ...Generator free pattern memory Defined 1 to 4 is set Parameter Numeric 00000 to 65535 a decimal number Unit Non term Restriction This cannot be set during the data pattern delivery but can be set in th...

Страница 28: ...it length of the Pattern Generator free pattern memory Defined 1 to 4 is inquired Response The delivery bit length of the Pattern Generator free pattern memory Defined 1 to 4 Restriction None PATTern...

Страница 29: ...length data Unit Non term Restriction This cannot be written during the data pattern delivery but can be written during idling Write the data after the data erasure PATTern DATA IDLE Numeric Character...

Страница 30: ...he data from the free pattern memory before writing data Also when changing written address data erase all the data from the memory by executing the ERASE command and then rewrite data from the first...

Страница 31: ...o bit end 0 1 0 0 1 1 0 1 1 0 1 0 0 1 0 0 0 1 0 0 1 1 0 1 1 0 1 0 0 1 0 1 0 0 1 1 0 1 1 0 1 0 0 1 0 PATTern DATA DEFined1 ERASe PATTern DATA DEFined1 00000 10110010 PATTern DATA DEFined1 00001 0010010...

Страница 32: ...24 Section 5 Remote Control by GPIB...

Страница 33: ...t delivery address DEF3 First delivery address DEF4 First delivery address IDLE First delivery address DEF1 data bit length DEF2 data bit length DEF3 data bit length DEF4 data bit length IDLE data bit...

Страница 34: ...26 Appendix Initialization...

Страница 35: ...Option 23 Pattern Generator Read this manual before using the equipment Keep this manual with the equipment MG3641A MG3642A Synthesized Signal Generator Operation Manual...

Страница 36: ...5 10 27 Minamiazabu Minato ku Tokyo 106 Japan Phone 81 3 3446 1111 Printed in Japan ANRITSU CORPORATION Document No M W1389AW 1 0...

Отзывы: