Software Operation and Overview
3-2 GUI Overview
MA25101A UG
PN: 10580-00476 Rev. B
3-3
IQ Acquisition Toolbar
The toolbar provides convenient controls for saving streaming data and interfacing with the analyzer.
Table 3-1.
IQ Acquisition Toolbar
Press the information button to 'connect' to the analyzer and show a summary of
the connected analyzer (see
Figure 3-2, “Analyzer Information Dialog”
indicates if a PCIe connection is available. If an analyzer is connected, the LED
button next to "Analyzer URL" will change color and the "Stream IQ Data button"
will be enabled. Refer to
“Status LED and Analyzer URL” on page 3-4
for
discussion of status and displayed messages.
Downloads the current analyzer settings to the
“IQ Acquisition Settings Panel”
.
Initiates IQ data streaming and saves the streaming data to the output folder (see
“IQ Acquisition Settings Panel” on page 3-4
). After it is pressed, the button
changes to
Stop IQ Stream
and it can be used to stop the stream. It is only
enabled if the analyzer is connected and the correct options are installed. The
button will be unavailable if a required option is missing or if a connection via the
selected streaming port is not established.
Opens Windows Explorer to the output folder location.
Deletes all supported streaming files (.dgz, .dgzm, .biq, .xdat, .xml, .xhdr) in the
specified output folder (see
“IQ Acquisition Settings Panel” on page 3-4
).
Opens a dialog box where the user can perform certain system performance
checks and optimize some system settings (see
)
Factory reset restores the analyzer to the factory default settings for all
measurement modes and system settings, including language and the display
and audio settings. Ethernet settings and user files are not affected. The analyzer
will automatically restart.
After the reset, the analyzer will need reconfiguring
and the IQ Acquisition software will need to be restarted.
Factory reset could
be useful to correct an erroneous analyzer setting that might affect the IQ stream
as it is the surest way to get the analyzer back into a known good state. A
message is displayed and the user can choose whether to continue with this
action.
Opens the IQ Auto-Capture panel. See
.
Opens the online help in the default browser.