5-6
POWER LEVEL
ACCURACY AND
FLATNESS TESTS
These tests verify that the power level accuracy and flatness of the
681XXC meet specifications. Table 5-3, beginning on page 5-19, con-
tains test records that you can copy and use to record test results for
these tests. Test records are provided for each 681XXC model configu-
ration.
Test Setup
Connect the equipment, shown in Figure 5-2, as
follows:
Step 1
Calibrate the Power Meter with the
Power Sensor.
NOTE
For
£
40 GHz models, use the
M A 2 4 7 4 A p o w e r s e n s o r ; f o r
>40 GHz models, use the MA2475A
power sensor.
Step 2
Connect the Power Sensor to the
RF OUT-
PUT
of the 681XXC.
Step 3
Connect the 681XXC rear panel
HORIZ
OUT
to the Oscilloscope CH.1 input (X in-
put).
NOTE
Before starting these procedures,
locate the test record in Table 5-3
for the particular 681XXC model
configuration being tested.
681XXC OM
5-13
OPERATION
POWER LEVEL ACCURACY
VERIFICATION
AND FLATNESS TESTS
R F
O U T
H O R I Z
O U T
6 8 1 X X C S I G N A L G E N E R A T O R
O S C I L L O S C O P E
P o w e r
S e n s o r
P O W E R M E T E R
C H 1 o r X I n p u t
Figure 5-2.
Equipment Setup for Power Level Accuracy and Flatness Tests
Содержание 680XXC
Страница 3: ......
Страница 11: ...Table of Contents Continued...
Страница 13: ...Figure 1 1 Series 681XXC Synthesized Signal Generator...
Страница 21: ......
Страница 37: ......
Страница 127: ...Table of Contents 4 1 INTRODUCTION 4 3 4 2 MENU MAP DESCRIPTION 4 3 Chapter 4 Local Operation Menu Maps...
Страница 128: ......
Страница 130: ......
Страница 131: ...LOCAL OPERATION SAMPLE MENU MAPS MENU MAP Figure 4 1 Sample Menu Map Annotated 681XXC OM 4 5...
Страница 146: ......
Страница 191: ......
Страница 206: ......