Test
Procedure
The following procedure tests both the coarse and
fine loops to verify the accuracy of the CW frequency
output.
Step 1
Set up the 681XXC as follows:
a. Reset the instrument by pressing
SYSTEM
, then
RESET
. Upon reset,
the CW Menu (below) is displayed.
b. Press
Edit F1
to open the current fre-
quency parameter for editing.
c. Set F1 to the first test frequency for
the model being tested (Table 5-2A is
the test record for standard models;
Table 5-2B for models with Option 11).
Step 2
Verify that the Frequency Counter read-
ing meets specifications (
±
100 Hz of the
value shown on the test record for stan-
dard models;
±
10 Hz for instruments with
Option 11).
Step 3
Record the Frequency Counter reading on
the test record (Table 5-2A or Table 5-2B).
NOTE
The Frequency Counter reading is
typically within
±
1 Hz. Differences
of a few Hertz can be caused by
noise or counter limitations. Differ-
ences of
³±
100 Hz (
³±
10 Hz for in-
struments with Option 11) indicate
a frequency synthesis problem.
Step 4
Set F1 to the next test frequency on the
test record and record the Frequency
Counter reading.
Step 5
Repeat step 4 until all frequencies listed
on the test record have been recorded.
5-6
681XXC OM
OPERATION
CW FREQUENCY
VERIFICATION
ACCURACY TEST
Содержание 680XXC
Страница 3: ......
Страница 11: ...Table of Contents Continued...
Страница 13: ...Figure 1 1 Series 681XXC Synthesized Signal Generator...
Страница 21: ......
Страница 37: ......
Страница 127: ...Table of Contents 4 1 INTRODUCTION 4 3 4 2 MENU MAP DESCRIPTION 4 3 Chapter 4 Local Operation Menu Maps...
Страница 128: ......
Страница 130: ......
Страница 131: ...LOCAL OPERATION SAMPLE MENU MAPS MENU MAP Figure 4 1 Sample Menu Map Annotated 681XXC OM 4 5...
Страница 146: ......
Страница 191: ......
Страница 206: ......