Test Records
A-2
PN: 10580-00195 Revision B
S312D MM
VNA Frequency Accuracy
Use the first row for the VNA Frequency Accuracy test. If the test fails, then use the second row for the
repeated test.
VNA Return Loss Verification
S312D
Firmware Revision:
________________
Operator:
________________
Date:
______________
Serial Number:
________________
Options:
___________________________________________________
Table A-1.
VNA Frequency Accuracy
Measured Value
Specifications
Hz
1000 MHz ± 75 kHz
Hz
1000 MHz ± 75 kHz
Table A-2.
VNA Return Loss Verification
Offset
Measured Min. Value
Min. Specification
Measured Max. Value
Max. Specification
20 dB
dB
≥
–21.7 dB
dB
≤
–18.3 dB
6 dB
dB
≥
–7.2 dB
dB
≤
–4.8 dB
Содержание Site Master S312D