Document No. 992040.  Issue 8 – 09.11.22                                                                                                                                            Page 5  

3.

 

Manoeuvring and Handling 

For safety, it is recommended that the accessory stand is moved by two people, and the 
stand should only be pushed in the directions indicated in Fig.1. 
 
When  loaded  with  accessories,  special  care  must  be  taken  when  manoeuvring  the 
accessory stand; ensure the stand is loaded correctly, as described in  Section 2, and the 
brakes are disengaged before moving off. 

 

 

WARNING:  Attempting  to  push  the  stand  incorrectly,  or  when  incorrectly 
loaded, could result in instability, tipping and injury to the user. 

 

WARNING: Extra care  should be taken when traversing ramps, thresholds and 
stray power cables. 

 

4.

 

Using the Brakes 

The brakes are located on each castor of the stand and are engaged by pressing down 
on the lever until it locks with a ‘click’. To disengage the brake press the lever back into 
the neutral position. 
 

 

WARNING:  To  prevent  injury  to  the  user  the  castor  brakes  must  be  engaged 
whilst loading and unloading accessories from the stand. 

 

5.

 

Cleaning and Disinfecting 

It  is  recommended  that  only  CE  marked  cleaning  products  are  used  in  the  cleaning  of 
this device.  Cleaning and Disinfection should be carried out by hand only. 
 
Clean the accessory stand with warm water and neutral detergent and dry the surfaces 
thoroughly  using  a  soft  cloth.    Suitable  disinfectants  are:  quaternary  ammonium 
compounds,  isopropyl  alcohol  &  chlorine  bleach  up  to  1%  (10,000  ppm).  Apply 
disinfectant by cloth, spray or disinfectant wipe.  Following disinfection, wash off the all 
surfaces with clean warm water and dry thoroughly using a soft cloth. 
 
The product will be adversely affected and its life expectancy reduced if the  following 
cautions are not observed; 

 

 

CAUTION: 

-

 

Do not steam clean or jet wash any area of this device. 

-

 

Do not soak or immerse the device. 

-

 

Do not use concentrated bleaching disinfectant solutions, organic solvent or 
abrasive powders in the cleaning or disinfection of this product. 

-

 

Dilute all disinfectants in accordance with the manufacturer’s guidelines. 

-

 

Disinfectant  products  are  corrosive  in  nature;  failure  to  properly  wash  and 
dry  the  product  surface  could  leave  a  corrosive  residue  which  may  cause 
damage to the product. 

 

6.

 

Product Maintenance 

The  accessory  stand  does  not  require  servicing.    However,  before  use,  visually  inspect 
the  stand  for  any  loose  or  damaged  parts  and  foreign  bodies  caught  in  the  castors. 
Every  six months we recommend the  stand is inspected by a competent person, i.e. an 
Anetic  Aid  service  engineer,  to  ensure  that  it  is  safe  to  use;  contact  Anetic  Aid  for 
further details. 
 
NOTE:  If  the  accessory  stand  is  damaged  or  faulty  it  must  be  taken  out  of  use  with 
immediate  effect  and  the  fault  reported  to  Anetic  Aid,  your  authorised  dealer  or 
maintenance  department.    The  stand  must  not  be  used  until  the  damage  or  fault  has 
been repaired. 
 
 

Содержание QA4

Страница 1: ...Document No 992040 Issue 8 09 11 22...

Страница 2: ...ironment for the storage and easy accessibility of operating theatre accessories and specific QA4 day surgery stand items CONTRAINDICATIONS The accessory stand should not be used as a walking aid The...

Страница 3: ...ssue 8 09 11 22 Page 2 The Storage Trough is suspended from the sidebar see fig 2 It is important that the accessory stand is cleaned and disinfected before it is put into use refer to Section 5 Clean...

Страница 4: ...Document No 992040 Issue 8 09 11 22 Page 3 2 Product Specifications...

Страница 5: ...ose WARNING Care must be taken when positioning accessories on the stand liability cannot be accepted for any damage to the equipment or injury to the user from improper use or overloading of the stan...

Страница 6: ...uitable disinfectants are quaternary ammonium compounds isopropyl alcohol chlorine bleach up to 1 10 000 ppm Apply disinfectant by cloth spray or disinfectant wipe Following disinfection wash off the...

Страница 7: ...Aid to confirm its receipt installation date and product details The warranty does not provide cover for breakage or failure due to tampering misuse neglect accidents modifications or shipping The wa...

Страница 8: ...t Full Width Articulating with EU Side Bar 21325 Headrest Dual articulating Head Positioner 21354 Leg Section Articulating with 105mm EU Side Bar 21355 Leg Section Articulating 21356 Leg Section Stand...

Отзывы: