DN:KD-926-SMSY01 V1.0                                                                    Operation Guide 

 

17/16 

Table 4 

For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING 

Recommended separation distances between 

portable and mobile RF communications equipment and the KD-926 

The KD-926 is intended for use in an electromagnetic environment in which radiated RF 

disturbances are controlled. The customer or the user of the KD-926 can help prevent 

electromagnetic interference by maintaining a minimum distance between portable and mobile 

RF communications equipment (transmitters) and the KD-926 as recommended below, 

according to the maximum output power of the communications equipment. 

Rated maximum 

output 

power of 

transmitter 

Separation distance according to frequency of transmitter 

150 kHz to 80 MHz 

 

 

80 MHz to 800 MHz 

 

 

800 MHz to 2,5 GHz 

 

 

0,01 

0.12 

0.12 

0.23 

0,1 

0.38 

0.38 

0.73 

1.2 

1.2 

2.3 

10 

3.8 

3.8 

7.3 

100 

12 

12 

23 

For transmitters rated at a maximum output power not listed above, the recommended 

separation distance d in meters (m) can be determined using the equation applicable to the 

frequency of the transmitter, where P is the maximum output power rating of the transmitter in 

watts (W) according to the transmitter manufacturer. 

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range 

applies. 

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is 

affected by absorption and reflection from structures, objects and people. 

 

P

d

2

.

1

P

d

2

.

1

P

d

3

.

2

Содержание KD-926

Страница 1: ...DN KD 926 SMSY01 V1 0 Operation Guide Operation Guide V1 0 Product project name Blood Pressure Monitor Model name KD 926 Project number Drafted by Date Reviewed by Date Approved by Date ...

Страница 2: ...Y LOADING AND AC ADAPTER LOADING 6 3 CLOCK AND DATE ADJUSTMENT 7 4 CONNECTING THE CUFF TO THE MONITOR 8 5 APPLYING THE CUFF 8 6 BODY POSTURE DURING MEASUREMENT 9 7 TAKING YOUR BLOOD PRESSURE READING 9 8 DISPLAYING STORED RESULTS 10 9 SYNCHRONIZING STORED RESULTS 10 10 DELETING MEASUREMENTS FROM THE MEMORY 11 11 ASSESSING HIGH BLOOD PRESSURE FOR ADULTS 11 12 TECHNICAL ALARM DESCRIPTION 12 13 TROUBL...

Страница 3: ... value usually appears in the daytime and lowest one usually at midnight Typically the value begins to increase at around 3 00 AM and reaches to highest level in the daytime while most people are awake and active Considering the above information it is recommended that you measure your blood pressure at approximately the same time each day Too frequent measurements may cause injury due to blood fl...

Страница 4: ...lood pressure and pulse rate The most recent 2 60 measurements can be stored in the memory with date and time stamp The Electronic Sphygmomanometer corresponds to the below standards IEC 60601 1 2005 A1 2012 E EN 60601 1 2006 A11 2011 Medical electrical equipment Part 1 General requirements for basic safety and essential performance IEC60601 1 2 2007 EN 60601 1 2 2007 AC 2010 Medical electrical eq...

Страница 5: ...5 Environmental humidity for storage and transport 85 RH 16 Environmental pressure 80kPa 105kPa 17 Battery life Approx 300 times 18 Wireless Connection Smart Bluetooth Frequency Band 2 400 2 4835GHz 19 All components belonging to the pressure measuring system including Pump Valve LCD Cuff Sensor Note These specifications are subject to change without notice NOTICE 1 Read all of the information in ...

Страница 6: ... you consult with your physician for accurate assessment There are 2 conditions under which the signal of IHB will be displayed 1 The coefficient of variation CV of pulse period 25 2 The difference of adjacent pulse period 0 14s and the number of such pulse takes more than 53 percentage of the total number of pulse 13 Please do not use the cuff other than supplied by the manufacturer otherwise it ...

Страница 7: ...n diabetes pre eclampsia enal diseases SETUP AND OPERATING PROCEDURES 1 DOWNLOAD THE FREE APP Prior to first use download and install the Andon App from the App Store iOS device or Google Play Android device Use keyword search terms mHealth 2 BATTERY LOADING AND AC ADAPTER LOADING a Open battery cover at the back of the monitor b Load four AA size batteries Please pay attention to polarity c Close...

Страница 8: ...de the time format will blink at first see picture3 1 The default time format is 24h Europe Version and the default clock and date is 2016 1 1 12 00 d Press the button START STOP repeatedly the year first usage default is 2016 range is 2016 2099 month day hour and minute will blink in turn see picture3 2 3 3 3 4 3 5 3 6 While the number is blinking press the button MEM to increase the number keep ...

Страница 9: ...air leakage during blood pressure measurements Avoid compression or restriction of the connection tubing during measurement which may cause inflation error or harmful injury due to continuous cuff pressure 5 APPLYING THE CUFF a Pulling the cuff end through the medal loop the cuff is packaged like this already turn it outward away from your body and tighten it and close the Velcro fastener b Place ...

Страница 10: ...ng your side with your palm upside c The cuff should be placed at the same level as your heart 7 TAKING YOUR BLOOD PRESSURE READING a After applying the cuff and your body is in a comfortable position press the START STOP button All display characters are shown for self test You can check the LCD display according to the right picture Please contact the service center if segment is missing See pic...

Страница 11: ...the last three results in this bank See picture 8 1 If no result stored LCD will show picture 8 2 c Press MEM button the LCD will display the average value of all results in the current user memory zone See picture 8 3 If no result stored in the current user memory zone LCD will display 0 for blood pressure and pulse rate See picture 8 4 d Press MEM button again the most recent result will be disp...

Страница 12: ... 8 5 c When Bluetooth is disconnected the monitor will turn off automatically after 1 minute of no operation You can also press the START STOP button to turn off the monitor manually 10 DELETING MEASUREMENTS FROM THE MEMORY When any result is displaying exclude average value displaying keeping on pressing button MEM for three seconds all results in the current bank will be deleted See picture 10 P...

Страница 13: ...eset in the factory and cannot be adjusted or inactivated This alarm condition is assigned as low priority according to IEC 60601 1 8 The technical alarm is non latching and need no reset The signal displayed on LCD will disappear automatically after about 8 seconds 13 TROUBLESHOOTING 1 PROBLEM POSSIBLE CAUSE SOLUTION LCD Display shows abnormal result The cuff position was not correct or it was no...

Страница 14: ... again after five minutes If the monitor is still abnormal please contact the local distributor or the factory LCD shows Er 6 More than 3 minutes with cuff pressure above 15 mmHg LCD shows Er 7 EEPROM accessing error LCD shows Er 8 Device parameter checking error LCD shows Er A Pressure sensor parameter error No response when you press button or load battery Incorrect operation or strong electroma...

Страница 15: ...fied technical personnel to repair those parts of equipment which are designated repairably can be supplied by us 9 The monitor can maintain the safety and performance characteristics for a minimum of 10 000 measurements or three years and the cuff integrity is maintained after 1 000 open close cycles of the closure 10 It is recommended the cuff should be disinfected 2 times every week if needed F...

Страница 16: ... This product complies with Industry Canada IC RSS 247 IC NOTICE This device complies with Industry Canada licence exempt RSS standard s Operation is sub ject to the following two conditions 1 this device may not cause interference and 2 this device must accept any interference including interference that may cause undesired operation of the device The Bluetooth word mark and logos are registered ...

Страница 17: ... 8kV 15kV air Radiated RF EM field IEC 61000 4 3 10V m 80MHz 2 7GHz 80 AM at 1kHz Proximity fields from RF wireless communications equipment IEC 61000 4 3 Refer to table 3 Rated power frequency magnetic fields IEC 61000 4 8 30A m 50Hz or 60Hz Table 3 Proximity fields from RF wireless communications equipment Test frequency MHz Band MHz Immunity test levels Professional healthcare facility environm...

Страница 18: ...ted maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be determined using the equation applicabl...

Страница 19: ...cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or rel...

Отзывы: