Hereby, [ANDON HEALTH CO.,LTD.] declares that the radio equipment type [KD-752] is in 

compliance with Directive 2014/53/EU. 

The full text of the EU declaration of conformity is available at the following internet address:

the internet address

 

 

ELECTROMAGNETIC COMPATIBILITY INFORMATION 

This product is applicable to the equipment and system requirements for the purpose of 

receiving radio frequency energy for the purpose of the work, Bluetooth receive bandwidth 2M. 

This product can also be used to include RF transmitter equipment and system requirements 

and emission frequency of 2.4GHz ISM band, Bluetooth modulation types:GFSK , effective 

radiated power: < 20dBm 

 

Table 1 - Emission 

Phenomenon 

Compliance 

Electromagnetic environment 

RF emissions 
 

CISPR 11 
Group 1, Class B 

Home healthcare environment 

Harmonic distortion 

IEC 61000-3-2 
Class A 

Home healthcare environment 

Voltage fluctuations 
and flicker 

IEC 61000-3-3 
Compliance 

Home healthcare environment 

 

Table 2 - Enclosure Port 

Phenomenon 

Basic EMC standard 

Immunity test levels 

Home Healthcare Environment 

Electrostatic 
Discharge 

IEC 61000-4-2 

±8 kV contact 
±2kV, ±4kV, ±8kV, ±15kV air 

Radiated RF EM field  IEC 61000-4-3 

10V/m 
80MHz-2.7GHz 
80% AM at 1kHz 

Proximity  fields  from 
RF 

wireless 

communications 
equipment 

IEC 61000-4-3 

Refer to table 3 

Rated 

power 

frequency  magnetic 
fields   

IEC 61000-4-8 

30A/m 
50Hz or 60Hz 

 

Table 3 

–    Proximity fields from RF wireless communications equipment 

Test frequency 

(MHz)

 

Band 

(MHz)

 

Immunity test levels 

Professional healthcare facility environment 

385 

380-390 

Pulse modulation 18Hz, 27V/m 

450 

430-470 

FM, 

±

5kHz deviation, 1kHz sine, 28V/m 

710 

704-787 

Pulse modulation 217Hz, 9V/m 

745 
780 

Содержание KD-752

Страница 1: ...DN KD 752 SMSY01 V1 0 Operation Guide History Revision NO Revision date Revision Description Versi on Drafter Approver 01 2018 10 25 Originated V1 0 Zhang RuoShang Zhang Fei...

Страница 2: ...D 752 SMSY01 V1 0 Operation Guide Operation Guide V1 0 Product project name Wireless Blood Pressure Monitor Model name KD 752 Project number 2017 007 XY13 Drafted by Date Reviewed by Date Approved by...

Страница 3: ...PROCEDURES 6 1 BATTERY LOADING 6 2 CLOCK AND DATE ADJUSTMENT 7 3 VOICE SETTING 8 4 CONNECTING THE CUFF TO THE MONITOR 9 5 APPLYING THE CUFF 9 6 BODY POSTURE DURING MEASUREMENT 10 7 TAKING YOUR BLOOD P...

Страница 4: ...value usually appears in the daytime and lowest one usually at midnight Typically the value begins to increase at around 3 00 AM and reaches to highest level in the daytime while most people are awake...

Страница 5: ...od pressure and pulse rate The most recent 4 60 measurements can be stored in the memory with date and time stamp The Electronic Sphygmomanometer corresponds to the below standards IEC 60601 1Edition...

Страница 6: ...m More than 60 incl 5 12 Environmental temperature for operation 10 40 50 104 13 Environmental humidity for operation 85 RH 14 Environmental temperature for storage and transport 20 50 4 122 15 Enviro...

Страница 7: ...please see part ELECTROMAGNETIC COMPATIBILITY INFORMATION It is suggested that the wireless blood pressure monitor be kept 10 meters away from other wireless devices such as WLAN unit cell phone micr...

Страница 8: ...2 Consult your physician before using the device for any of the following conditions common arrhythmias such as atrial or ventricular premature beats or atrial fibrillation arterial sclerosis poor per...

Страница 9: ...OCK AND DATE ADJUSTMENT a Once you install the battery it will enter Clock Mode and LCD will display time and date See picture 2 1 The monitor will enter Standby Mode after 10 seconds with no operatio...

Страница 10: ...00 12 00 PM 1 00 1 00 AM 13 00 1 00 PM 2 00 2 00 AM 14 00 2 00 PM 3 00 3 00 AM 15 00 3 00 PM 4 00 4 00 AM 16 00 4 00 PM 5 00 5 00 AM 17 00 5 00 PM 6 00 6 00 AM 18 00 6 00 PM 7 00 7 00 AM 19 00 7 00 PM...

Страница 11: ...ttached to the monitor when it is packaged Should the cuff become unattached align the two plugs and four brackets of the cuff with the plug sockets and bracket sockets of the monitor and press the cu...

Страница 12: ...more reliable than that from only one measurement Slide the standard multiple switch on the right side of the monitor toward to 1 to select standard mode and toward to MULT to select multiple mode be...

Страница 13: ...peration Alternatively you can press the START button to turn off the monitor manually e The measurement result will be shown after multiple measurements The individual results will not be displayed f...

Страница 14: ...no result which is measured from 18 o clock to 20 o clock in last 7 days stored in the current user memory zone the LCD will display 0 See picture 8 9 f Press MEM button again the most recent result w...

Страница 15: ...o operation You can also press the START button to turn off the monitor manually 10 DELETING MEASUREMENTS FROM THE MEMORY When any result is displaying exclude average value displaying keeping on pres...

Страница 16: ...d Apply the cuff correctly and try again Body posture was not correct during testing Review the BODY POSTURE DURING MEASUREMENT sections of the instructions and re test Speaking arm or body movement a...

Страница 17: ...sure sensor parameter error No response when you press button or load battery Incorrect operation or strong electromagnetic interference Take out batteries for five minutes and then reinstall all batt...

Страница 18: ...soft cloth squeezed well after moistened with water diluted disinfectant alcohol or diluted detergent 10 No component can be maintained by user in the monitor The circuit diagrams component part lists...

Страница 19: ...INFORMATION Only charge the cost of components and transport SERVICE CENTER ANDON HEALTH CO LTD No 3 Jinping Street YaAn Road Nankai District Tianjin 300190 China Tel 86 22 60526081 iHealthLabs Europ...

Страница 20: ...magnetic environment RF emissions CISPR 11 Group 1 Class B Home healthcare environment Harmonic distortion IEC 61000 3 2 Class A Home healthcare environment Voltage fluctuations and flicker IEC 61000...

Страница 21: ...ide 19 19 810 800 960 Pulse modulation 18Hz 28V m 870 930 1720 1700 1990 Pulse modulation 217Hz 28V m 1845 1970 2450 2400 2570 Pulse modulation 217Hz 28V m 5240 5100 5800 Pulse modulation 217Hz 9V m 5...

Отзывы: