Hereby, [ANDON HEALTH CO.,LTD.] declares that the radio equipment type [KD-752] is in
compliance with Directive 2014/53/EU.
The full text of the EU declaration of conformity is available at the following internet address:
【
the internet address
】
ELECTROMAGNETIC COMPATIBILITY INFORMATION
This product is applicable to the equipment and system requirements for the purpose of
receiving radio frequency energy for the purpose of the work, Bluetooth receive bandwidth 2M.
This product can also be used to include RF transmitter equipment and system requirements
and emission frequency of 2.4GHz ISM band, Bluetooth modulation types:GFSK , effective
radiated power: < 20dBm
Table 1 - Emission
Phenomenon
Compliance
Electromagnetic environment
RF emissions
CISPR 11
Group 1, Class B
Home healthcare environment
Harmonic distortion
IEC 61000-3-2
Class A
Home healthcare environment
Voltage fluctuations
and flicker
IEC 61000-3-3
Compliance
Home healthcare environment
Table 2 - Enclosure Port
Phenomenon
Basic EMC standard
Immunity test levels
Home Healthcare Environment
Electrostatic
Discharge
IEC 61000-4-2
±8 kV contact
±2kV, ±4kV, ±8kV, ±15kV air
Radiated RF EM field IEC 61000-4-3
10V/m
80MHz-2.7GHz
80% AM at 1kHz
Proximity fields from
RF
wireless
communications
equipment
IEC 61000-4-3
Refer to table 3
Rated
power
frequency magnetic
fields
IEC 61000-4-8
30A/m
50Hz or 60Hz
Table 3
– Proximity fields from RF wireless communications equipment
Test frequency
(MHz)
Band
(MHz)
Immunity test levels
Professional healthcare facility environment
385
380-390
Pulse modulation 18Hz, 27V/m
450
430-470
FM,
±
5kHz deviation, 1kHz sine, 28V/m
710
704-787
Pulse modulation 217Hz, 9V/m
745
780