DN:KD-5985-SMSY01 V3.0 Operation Guide
Template version:1
20/22
that the KD-5985
be powered from
an uninterruptible
power supply or a
battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields
should be at
levels
characteristic of a
typical location in
a typical
commercial or
hospital
environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
Table 3
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer’s declaration - electromagnetic immunity
The KD-5985 is intended for use in the electromagnetic environment specified below. The
customer or the user of the KD-5985 should assure that it is used in such an environment.
IMMUNITY
test
IEC 60601test
level
Compliance
level
Electromagnetic environment -
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms 150 kHz to
80 MHz
3 V/m 80 MHz to
2,5 GHz
3 Vrms
3 V/m
Portable and mobile RF
communications equipment should
be used no closer to any part of the
KD-5985, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation
distance
:
80 MHz to 800 MHz
800 MHz to 2,5 GHz
P
d
2
.
1
P
d
2
.
1
P
d
3
.
2