background image

P

AGE

  3

Section 1 - General

Section 1 - General

Section 1 - General

Section 1 - General

Section 1 - General

1.1  MOUNTING

1.1  MOUNTING

1.1  MOUNTING

1.1  MOUNTING

1.1  MOUNTING

The Model HC1-series electrodeless conductivity sensor measures the conductivity of solutions

from 0-200 up to 0-2,000,000 microSiemens/cm.
The sensor is constructed of 3A approved materials and has a sanitary 2" tri-clamp process

connection.  The sensor may be installed in a special 2" x 2" sanitary tee (available from

Anderson), or in a standard 2 1/2" x 2", or 3" x 2" short outlet reducing tee.

1.2  WETTED MATERIALS

1.2  WETTED MATERIALS

1.2  WETTED MATERIALS

1.2  WETTED MATERIALS

1.2  WETTED MATERIALS

The sensor is constructed of 

PVDF or 

PFA Teflon

 depending on option choice

.

1.3  OPERATING PRECAUTIONS

1.3  OPERATING PRECAUTIONS

1.3  OPERATING PRECAUTIONS

1.3  OPERATING PRECAUTIONS

1.3  OPERATING PRECAUTIONS

Always consider the temperature/pressure ratings of the mounting hardware used to install the

sensor. The sensor and hardware combine to become an integrated system. The hardware

material usually limits the system’s temperature/pressure rating. Refer to Section 2 for complete

specifications.

Section 2 - Specifications

Section 2 - Specifications

Section 2 - Specifications

Section 2 - Specifications

Section 2 - Specifications

Wetted Materials

PVDF or 

PFA Teflon®

Operating  Temperature  Range

14 to 3

47

°F (-10 to 

175

°C)

Measuring  Range

From 0-200 to 0-2,000,000 microSiemens/cm

Temperature  Compensator

Pt 1000 RTD

Sensor  Cable:

5 conductor (plus two isolated shields) cable with Teflon-

coated jacket; rated to 

175

°C (3

47

°F); 20 ft. (6 m) long

Pressure/Temperature  Limits:

          

200 psi at 3

47

°F (13.8 bar at 

175

°C)

* Ratings for above sensor are based upon water service.  More severe service may require a

correction factor.

Teflon

® 

is a registered DuPont Co. Trademark.

Section 3 - Location Requirements

Section 3 - Location Requirements

Section 3 - Location Requirements

Section 3 - Location Requirements

Section 3 - Location Requirements

Locate the sensor as close as possible to the instrument. Depending on the instrument model

and full-scale range, the maximum allowable distance between sensor and instrument varies

(refer to instrument manual for distances).

NOTE: When indirectly wiring the sensor to the instrument with interconnect cable and a

junction box, use only an Anderson specified cable. When using a different cable, it
must have equivalent construction: five conductors and two separate isolated shields
— one shielding the signal, and one shielding the overall cable. These specific cable
characteristics protect the measurement signal from electromagnetic interference.
Using a cable with different construction may interfere with the measuring system’s
performance.

Содержание HC1 Series

Страница 1: ..._______________________ Instrument Serial Number ____________________________ Form Number AIC2032 June 2003 Rev 2 13 Model HC1 Series Model HC1 Series Model HC1 Series Model HC1 Series Model HC1 Series Electrodless Conductivity Sensors Electrodless Conductivity Sensors Electrodless Conductivity Sensors Electrodless Conductivity Sensors Electrodless Conductivity Sensors ...

Страница 2: ... STANDARD TERMINATION PROCEDURE EXCEPT CE 8 5 3 STANDARD TERMINATION PROCEDURE EXCEPT CE 8 5 4 INTERCONNECT CABLE DETAILS 9 5 4 INTERCONNECT CABLE DETAILS 9 5 4 INTERCONNECT CABLE DETAILS 9 5 4 INTERCONNECT CABLE DETAILS 9 5 4 INTERCONNECT CABLE DETAILS 9 Section 6 T Section 6 T Section 6 T Section 6 T Section 6 Troubleshooting roubleshooting roubleshooting roubleshooting roubleshooting 10 10 10 1...

Страница 3: ...ications Section 2 Specifications Wetted Materials PVDF or PFA Teflon Operating Temperature Range 14 to 347 F 10 to 175 C Measuring Range From 0 200 to 0 2 000 000 microSiemens cm Temperature Compensator Pt 1000 RTD Sensor Cable 5 conductor plus two isolated shields cable with Teflon coated jacket rated to 175 C 347 F 20 ft 6 m long Pressure Temperature Limits 200 psi at 347 F 13 8 bar at 175 C Ra...

Страница 4: ... junction box that has a terminal strip onto a flat surface Make sure the junction box cover can be removed after installation 3 Before installing the sensor into the process line or vessel calibrate the measuring system with conductivity reference solution using the procedure in the instrument operating manual 4 After calibration mount the sensor into the sanitary tee or ferrule A Properly seat t...

Страница 5: ...PAGE 5 FIGURE 4 1 Sanitary Mounting Details 2 67 8 54 1 91 2 67 9 08 1 91 ...

Страница 6: ...tions except those using instruments with the European Community CE label The sensor cable is terminated with five colored wires red green yellow white and blue and one shield wire signal shield wire and cable shield wire are electrically common CE Termination Available by special order for applications using instruments with the European Community CE label The sensor cable is terminated with five...

Страница 7: ...OCEDURE TO TERMINATE THE SENSOR OR INTERCONNECT CABLE UNLESS THE MEASURING INSTRUMENT HAS THE EUROPEAN COMMUNITY CE LABEL IN THIS CASE USE THE CE TERMINATION PROCEDURE DESCRIBED IN SECTION 5 3 1 Refer to Figure 5 1 and strip back 3 inches 76 mm of the outer cable jacket outer foil wrap cellophane wrap inner foil wrap and unused black wire This exposes the signal shield wire the cable shield wire a...

Страница 8: ...il wrap This exposes the signal shield wire cable shield wire and cellophane wrapped wire bundle 2 Fold back the cable shield wire Using shrink wrap or tape equally cover the outer cable jacket and the exposed cable shield wire cellophane wrapped wire bundle 3 Carefully clip back the cellophane wrap inner foil wrap and unused black wire up to the end of the shrink wrap or tape applied in step 2 Th...

Страница 9: ...pecific cable characteristics protect the measurement signal from electromagnetic interference Using a cable with different construction may interfere with the measuring system s ability to operate properly If the application requires interconnect beyond the standard 20 length the use of a Interconnect Cable Junction box is required to ensure a quality connection 1 Mount the junction box on a flat...

Страница 10: ...an 5 ohms 5 Place the ohmmeter to its highest range and measure the resistance between the shield and white wires The reading should indicate infinity open circuit If any one of the resistance checks in the previous steps is incorrect the sensor is probably defective Contact Anderson Instrument Company for details on arrang ing a sensor replacement If the resistance checks are correct the sensor m...

Страница 11: ...ntial damages special damages or any other damages costs or expenses excepting only the cost or expense of repair or replacement as described above Products must be installed and maintained in accordance with Anderson instructions Users are responsible for the suitability of the products to their application There is no warranty against damage resulting from corrosion misapplication improper speci...

Отзывы: