ANDERSON-NEGELE MPI-200 Скачать руководство пользователя страница 6

C O N T R O L S

6

Printing/exporting the parameter list

Selection options/input fi elds

Printing and/or exporting parameters

Depending on the activation of a button in the button bar (see also page 3) or using the submenus described below, the 

parameter settings can be printed out in a list, duplicates of the list can be created or the parameter data of a sensor can be 

saved and copied to other identical sensors.

Submenu items

1:  Print the parameter list directly

2:  Save parameter data set on the PC

3:  Load the saved parameter data set 

from the PC to the sensor

Calling up submenus

1.
2.
3.

Selection options

In the next step in each submenu, you 

can select which parameter families 

should be saved or printed (1.).

Additional input fi elds are available (2.) 

to assign a parameter list or a data set 

to a particular sensor/unit.

The serial number is a mandatory fi eld; 

all other parameters are optional. 

These optional fi elds show the informa-

tion of the sensor that was processed 

last. Unless this information is changed 

or deleted, it will also be printed out 

and saved.

After the language and the printer are 

selected in the option fi eld (3.), the 

process can be completed by clicking 

the “OK” button.

1.

2.

3.

Export function of the parameter list

After the print menu is opened and the corresponding data are entered, a paper printout can be created.

To make sure you always have a copy of the parameter settings available in the form of a list (e.g. for quality assurance), it is 

recommended that you save the parameter list in .xml format in addition to printing it out.

Export/import of the parameter list

The import and export function can be used to load parameter settings onto another, identical device. This function can be 

used to store parameter data sets externally on the PC and to load them again when needed.

Information

The exported xml fi le can be saved locally on the PC or on a server and can be distributed/sent by e-mail or USB stick. 

This does not require special database or programming knowledge. The data is saved in the Microsoft Windows direc-

tory structure. The default fi le name depends on the entries in the serial number and TAG number fi elds. However, it can 

be changed or expanded manually if required. This import function can be used to load the xml fi le and upload it onto a 

sensor.

Содержание MPI-200

Страница 1: ...her requirement CD drive for program installation Access rights Administrator rights for installation Installing the software 1 Insert the CD included with the MPI 200 into the CD drive 2 Double click...

Страница 2: ...external M12 connected A supply from the PC would otherwise be superimposed on the 4 20 mA output signal and lead to an incorrect measur ing signal Electrical connection Power supply Connecting the se...

Страница 3: ...t Reconnect Establishing a connection with a sensor reading in the parameter structure again manually Mode Mode setting Monitoring Process values and service information can be viewed but not changed...

Страница 4: ...ed e g 4 mA for a mA setpoint Save parameter The manually created setting is adopted and the display returns to the main menu Discard parameter The setting is not adopted and the display returns to th...

Страница 5: ...at can only be set to xed values By pressing the arrow buttons Only values that are greater than the cur rently displayed value are possible Only values less than the currently displayed value are pos...

Страница 6: ...out and saved After the language and the printer are selected in the option field 3 the process can be completed by clicking the OK button 1 2 3 Export function of the parameter list After the print m...

Страница 7: ...supply MPI 200 parameter set to PC Set the appropriate power supply Error message when the software is started Is NET 4 0 or higher installed on the PC Install NET from the supplied CD path NET4 0 Er...

Страница 8: ...te Note the cleaning information To avoid damage of the equipment use suitable trans port packaging only Transport storage Do not store outside Store in an area that is dry and dust free Do not expose...

Отзывы: