Technical Data
Type UM-211
Measurement method
Oscillometric measurement
Measurement range
Pressure: 0 - 299 mmHg
Systolic pressure: 60 - 279 mmHg
Diastolic pressure: 40 - 200 mmHg
Pulse:
40 - 200 beats / minute
Measurement accuracy
Pressure: ±3 mmHg
Pulse:
±5%
Temperature unit
°C or °F
Temperature accuracy
±2.5°C (+5°C to +40°C)
Power supply
Built-in 3.6V battery (UM-211-20) or
AC adapter (TB-268)
Number of measurements
Approx. 300 measurements, when built-in battery
is used, with pressure value of 180 mmHg at room
temperature of 23°C
Classification Internally
powered ME equipment (Supplied by
batteries) /
Class II (Supplied by adapter)
Continuous operation mode
Clinical test
According to ISO8106
0
-2 2013
EMC
IEC 60601-1-2: 2007
Memory Last
99
measurements
Operating condition
+5°C to +40°C / 10%RH to 85%RH (Not condensed)
800 hPa to 1060 hPa
Transport / Storage conditions -20°C to +60°C / 10%RH to 95%RH (Not condensed)
700 hPa to 1060 hPa
Dimensions
Approx. 120 [W] x 200 [H] x 140 [D] mm
Weight
Approx. 550 g, excluding the battery
Applied part
Cuff Type BF
Useful life
Device: 5 years
Cuff: 2 years
AC adapter: 5 years